A&D HV-G Series instruction manual Preparation Setting Parameters

Page 27

Function table

Meaning & Purpose

f8 4

When the display is a stable value, the scale accumulates the data

 

automatically. The next accumulation can be performed after the

 

display returns to near-zero.

 

Used for recording the number and mass of articles removed from the

 

pan. (Put the articles on the pan. Press the

 

switch at each

 

TARE

 

removal.)

 

 

 

 

 

 

f8 5

At each finish of the full/dribble batch function, the scale accumulates

 

the data automatically.

 

Used for packaging articles like a powder, it is used for recording the

 

bag number and total mass.

 

 

 

 

10.1 Preparation (Setting Parameters)

Step

1

Turn off the display.

 

 

Press the

 

 

 

switch while the

 

 

switch is

 

 

ON/OFF

ZERO

 

 

pressed and held. The function table is displayed.

Step

2

Press the

 

switch to display an item of the

ENTER

 

 

accumulation function ( f8 ).

Step

3

Select a parameter of the accumulation function ( f8 1

 

 

~ f8 4 ) with the ^ switch.

Step

4

Press the

 

 

switch to store the new parameter.

 

ENTER

 

 

Press the F switch and the

 

 

switch to exit from

 

 

ENTER

 

 

the function table. Then the scale displays the weighing

 

 

mode.

HV-G Instruction Manual Page 26

Image 27
Contents HV-G Series Contents Page Compliance with FCC rules ComplianceOutline and Features Unpacking Type Products Accessories Accessories and Options listOptions List Order code or option name AccessoriesPrecautions for Installing the Scale Precautions for Operating the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names OFF Display and SymbolsPage Switches Page Basic Operation Turning the Scale On/Off & Basic WeighingMemo Type L with Batteries Semi-Automatic Tare Input by Weighing Tare and Net DisplayProducts Weighing Range Weighing RangeWeighing Range Operation and PerformanceMode Switch Changing Unit & Mode Storing a Unit Mass Counting ModeCounting the number of articles Storing a 100% Mass Percentage ModeReading Percentage Operation and Switches Accumulation FunctionFunction table Meaning & Purpose Preparation Setting Parameters Example Operation and Performance ExamplesFunction table Meaning and Purpose Comparator FunctionComparator Sign Comparison ConditionAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Comparator Full/Dribble batch sub-functionHold Parameter and display an item of the full/dribble Switch to display the HI Function Table Meaning and purpose Simple batch function Simple Batch FunctionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Gravity Acceleration Correction Complete Calibration ProcedureCalibration of the Zero Point Function Table Procedure for Setting Parameters15.2 Parameter List Function or full/dribble batch function Display Meaning and purposePrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Format Data FormatFormat Preparation and Performance Examples Stream ModeCommand Mode Cancel of tareTare Upper limit value Preset tareFinal value Preparation Setting Parameters Pin connections Pin No Signal name Direction Description Options17.1 RS-232C / Relay output / Buzzer OP-03 17.2 RS-422 / RS-485 / Relay output OP-04 Pin No Signal name Direction DescriptionInstalling OP-04 Communication Format Lower limit value ZeroFinal value Reset of Accumulation data and numberAccumulation data output Accumulation count outputSpecification Internal Printer for Type V OP-06Consumables Installation DisassemblyPerformance Check HV-G series SpecificationsDimensions Page This situation Confirm these items MaintenanceCheck Points Before Calling Maintenance Repair