14.2.3 Calibration of the Zero Point

Step 8 Press and hold the CAL switch to enter the calibration mode after displaying normal weighing for 30 minutes.

Then the Cal 0 is displayed.

Step 9 With nothing on the pan, press the ENTER switch while the sta- ble mark is displayed. The scale stores the current condition as the zero point. .

Step 10 The scale displays 5pn1 for several seconds. If finishing the calibration mode at this stage (Zero calibration only), proceed to step 14.

14.2.4 Span Calibration

Step 11 Set the value of the calibration mass using the following switches. (This initial value is according to

each product.)

^

switch

Selecting the value of

a

digit.

 

 

switch

Selecting a digit.

<

Step 12 Place the mass’ selected at step 11’ onto the pan and press the ENTER switch while the stable mark is displayed. The scale then calculates the span and stores it.

Step 13 The scale displays end at the finish.

Remove the mass from the pan.

Step 14 Press the CAL switch to return to the normal weighing mode.

HV-G Instruction Manual Page 42

Image 43
Contents HV-G Series Contents Page Compliance with FCC rules ComplianceOutline and Features Unpacking Type Products Accessories Accessories and Options listOptions List Order code or option name AccessoriesPrecautions for Operating the Scale Precautions for Installing the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names OFF Display and SymbolsPage Switches Page Turning the Scale On/Off & Basic Weighing Basic OperationMemo Type L with Batteries Semi-Automatic Tare Input by Weighing Tare and Net DisplayProducts Weighing Range Weighing RangeWeighing Range Operation and PerformanceMode Switch Changing Unit & Mode Storing a Unit Mass Counting ModeCounting the number of articles Storing a 100% Mass Percentage ModeReading Percentage Accumulation Function Operation and SwitchesFunction table Meaning & Purpose Preparation Setting Parameters Example Operation and Performance ExamplesFunction table Meaning and Purpose Comparator FunctionComparator Sign Comparison ConditionAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Full/Dribble batch sub-function ComparatorHold Parameter and display an item of the full/dribble Switch to display the HI Function Table Meaning and purpose Simple batch function Simple Batch FunctionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Gravity Acceleration Correction Complete Calibration ProcedureCalibration of the Zero Point Function Table Procedure for Setting Parameters15.2 Parameter List Function or full/dribble batch function Display Meaning and purposePrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Format Data FormatFormat Preparation and Performance Examples Stream ModeCancel of tare Command ModeTare Upper limit value Preset tareFinal value Preparation Setting Parameters Options Pin connections Pin No Signal name Direction Description17.1 RS-232C / Relay output / Buzzer OP-03 17.2 RS-422 / RS-485 / Relay output OP-04 Pin No Signal name Direction DescriptionInstalling OP-04 Communication Format Lower limit value ZeroFinal value Reset of Accumulation data and numberAccumulation data output Accumulation count outputInternal Printer for Type V OP-06 SpecificationConsumables Installation DisassemblyPerformance Check HV-G series SpecificationsDimensions Page This situation Confirm these items MaintenanceCheck Points Before Calling Maintenance Repair