14.2 The Complete Calibration Procedure

14.2.1 Gravity Acceleration Correction

Step1

Turn on the display.

 

Open the rear cover of the display unit.

 

Inside the display unit, press and hold the

 

 

switch to enter the calibration mode.

 

CAL

 

Then

Cal 0

is displayed.

Step2

Press the

 

 

 

switch to enter the gravity

ZERO

 

acceleration

 

correction mode.

Step3

Set your local gravity acceleration using

 

the following switches.

 

^ switch

Selecting the value of a digit.

 

< switch

Selecting a digit.

Step4

Press the

 

 

 

 

switch to store the new

ENTER

 

value.

 

 

 

 

 

Step5

Press the

 

 

switch again. Then the

CAL

 

scale returns to the normal weighing

 

mode.

 

 

 

 

 

14.2.2 Preparation

Step

6

Confirm the environmental conditions as follows:

 

 

Maintain a constant temperature and a stable power supply.

 

 

Install the scale on a solid floor or bench where there is no draft,

 

 

vibration, strong magnetic fields or direct sunlight.

 

 

Consider section “4. Caution”.

Step

7

Display normal weighing for at least 30 minutes to warm up the scale.

HV-G Instruction Manual Page 41

Image 42
Contents HV-G Series Contents Page Compliance Compliance with FCC rulesOutline and Features Unpacking Accessories and Options list Type Products AccessoriesOrder code or option name Accessories Options ListPrecautions for Installing the Scale Precautions for Operating the ScalePrecautions for Storing the Scale Installing the Scale Installing the Batteries for Type L Names Display and Symbols OFFPage Switches Page Basic Operation Turning the Scale On/Off & Basic WeighingMemo Type L with Batteries Tare and Net Display Semi-Automatic Tare Input by WeighingOperation and Performance Weighing RangeWeighing Range Products Weighing RangeMode Switch Changing Unit & Mode Counting Mode Storing a Unit MassCounting the number of articles Percentage Mode Storing a 100% MassReading Percentage Operation and Switches Accumulation FunctionFunction table Meaning & Purpose Preparation Setting Parameters Operation and Performance Examples ExampleComparison Condition Comparator FunctionComparator Sign Function table Meaning and PurposeAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Comparator Full/Dribble batch sub-functionHold Parameter and display an item of the full/dribble Switch to display the HI Simple Batch Function Function Table Meaning and purpose Simple batch functionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Complete Calibration Procedure Gravity Acceleration CorrectionCalibration of the Zero Point Procedure for Setting Parameters Function Table15.2 Parameter List Display Meaning and purpose Function or full/dribble batch functionPrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Data Format FormatFormat Stream Mode Preparation and Performance ExamplesCommand Mode Cancel of tareTare Preset tare Upper limit valueFinal value Preparation Setting Parameters Pin connections Pin No Signal name Direction Description Options17.1 RS-232C / Relay output / Buzzer OP-03 Pin No Signal name Direction Description 17.2 RS-422 / RS-485 / Relay output OP-04Installing OP-04 Communication Format Zero Lower limit valueAccumulation count output Reset of Accumulation data and numberAccumulation data output Final valueSpecification Internal Printer for Type V OP-06Consumables Disassembly InstallationPerformance Check Specifications HV-G seriesDimensions Page Repair MaintenanceCheck Points Before Calling Maintenance This situation Confirm these items