A&D HV-G Series instruction manual Function Table, Procedure for Setting Parameters

Page 44

15 The Function Table

The function table is provided to store and reference items that determine the performance of the scale. Each item has a parameter.

The parameters are maintained even without power applied.

15.1 The Procedure for Setting Parameters

Step 1 Turn off the display.

Press the ON/OFF switch while the ZERO switch is pressed and held. The function table is displayed.

Step 2 Set parameters for each item using the following

switches.

 

^ switch

Selecting the parameter of an item.

 

 

Selecting the value of a digit at f16.

< switch

Selecting a digit at f16..

F switch

Proceeding to the end of the table

 

 

without storing the parameter.

 

switch

Storing a parameter for the current

ENTER

 

 

item and proceeding to the next item.

 

 

Returning to normal mode from the

 

 

end of the table.

Step 3 Press the ENTER switch to return to the normal mode, when end is displayed.

Note Pressing the ENTER

stored in the scale.

switch on step 2, the parameter is

HV-G Instruction Manual Page 43

Image 44
Contents HV-G Series Contents Page Compliance Compliance with FCC rulesOutline and Features Unpacking Accessories and Options list Type Products AccessoriesOrder code or option name Accessories Options ListPrecautions for Storing the Scale Precautions for Installing the ScalePrecautions for Operating the Scale Installing the Scale Installing the Batteries for Type L Names Display and Symbols OFFPage Switches Page Memo Basic OperationTurning the Scale On/Off & Basic Weighing Type L with Batteries Tare and Net Display Semi-Automatic Tare Input by WeighingWeighing Range Weighing RangeOperation and Performance Products Weighing RangeMode Switch Changing Unit & Mode Counting Mode Storing a Unit MassCounting the number of articles Percentage Mode Storing a 100% MassReading Percentage Function table Meaning & Purpose Operation and SwitchesAccumulation Function Preparation Setting Parameters Operation and Performance Examples ExampleComparator Function Comparator SignComparison Condition Function table Meaning and PurposeAccumulation function f6 Step If one Has Been selected, press Case Full/Dribble Batch Function Hold ComparatorFull/Dribble batch sub-function Parameter and display an item of the full/dribble Switch to display the HI Simple Batch Function Function Table Meaning and purpose Simple batch functionPreparation Setting Parameters Operation and Performance Examples Calibration Adjusting the Scale Gravity Acceleration Table Complete Calibration Procedure Gravity Acceleration CorrectionCalibration of the Zero Point Procedure for Setting Parameters Function Table15.2 Parameter List Display Meaning and purpose Function or full/dribble batch functionPrinting condition when using option OP-06 for type Function to hold the current display value. When 16 RS-232C Serial Data Interface Data Format FormatFormat Stream Mode Preparation and Performance ExamplesTare Command ModeCancel of tare Preset tare Upper limit valueFinal value Preparation Setting Parameters 17.1 RS-232C / Relay output / Buzzer OP-03 Pin connections Pin No Signal name Direction DescriptionOptions Pin No Signal name Direction Description 17.2 RS-422 / RS-485 / Relay output OP-04Installing OP-04 Communication Format Zero Lower limit valueReset of Accumulation data and number Accumulation data outputAccumulation count output Final valueConsumables SpecificationInternal Printer for Type V OP-06 Disassembly InstallationPerformance Check Specifications HV-G seriesDimensions Page Maintenance Check Points Before Calling MaintenanceRepair This situation Confirm these items