How to Disinfect
The
be disinfected with benzalkonium chloride.
1 | Make 0.1% dilute solution using benzalkonium chloride, and soak the |
| Medication Bottle and Mesh Cap in the solution for 10 minutes. |
2 | Take the Medication Bottle and Mesh Cap out of solution |
| to rinse it with water. |
3 | Wipe off the remaining water for faster drying. |
How to Replace the Mesh Cap
You can usually use the Mesh Cap for approximately 1 year (according to the test results of OMRON). However, this may vary depending on medication type and individual use and care. In some cases, the unit may not nebulize or may nebulize weakly even after disinfection is done.
If the unit does not nebulize or nebulizes poorly even after cleaning the Mesh Cap,
purchase a new Mesh Cap from your nearest Omron dealer or where you purchased your unit from. The part number is
1Open the Mesh Cap.
Lift up the Mesh Cap by strongly applying your thumb on the brim.
4 If using commercial cleaners, follow manufacturers instructions. |
Caution
Disinfect | 5 | Do not use disinfection solution containing sodium hypochlorite. | |
The following parts can be disinfected: | |||
|
| • The unit may not be able to be used after getting covered with rust. | |
to |
| • | Medication Bottle |
How |
| ||
| • | Mesh Cap | |
|
| ||
|
| ||
Unit |
| • | Mask |
the |
| • | Mouthpiece |
DisinfecttoHow |
| ||
| • | Mouthpiece Adapter | |
|
|
2Remove the Mesh Cap from the Medication Bottle.
3Install a new Mesh Cap.
Insert a new Mesh Cap vertically.
Lift up the Mesh Cap vertically.
Close it securely.
How to Disinfect the Unit How to Replace the Mesh Cap
18 | 19 |