R-4106.A
2. Binocular Body
1)lf in
r;, \\ / | (/( |
|
|
|
(+) |
|
Fig. 37
2)After the adjustment oi :§ has been completed, check at ali times for the correct adjust· ments mentioned in
2._ 7. Finally, lóck prisms @ and @ in position using adhesive.
t:.... . e
1)lnsert the parfocality adjusting mirror (tool) into photo·tube sleeve @.
2)Setting the autocollimator to infinity, perform infinity checking, using the above mentioned
mirror.
3)If not adjusted to infinity, tum @ in relaiion to
4) Apply screw lock to G) and @), and leave them in th~'ir adjusted positions until the lock sets, to avoid parfocality derangement caused by screw slack.