DISASSEMBLY PROCEDURE | 3. UW·tube 4. ER·tube |
|
|
3.UW·tube
Release screw ~ xJ, and separate the binocular body from the depression pnsm unit.
1)Photo·tube
For disassembly, refer to Fig. 7, taking the cautions as below:
•When taking off Photo·tube sleeve e±) out of interna! tube G), mark off a line on the sleeve ©to regain the original position.
2)To disassemble the binocular body, refer to Fig. 9 and 1O.
3)To disassemble the depression prism unit, refer to Fig. 8, taking the cautions as below:
•Do not remove s1multaneously both the dovetail grooves (!I) from bottom plate @, because either is to be used as fiducial, when reassembling.
4.ER·tube
Release screws @ x J and separa te the binocular body from the depression prism unit.
1)Binocular body
For disassembly, follow the sarne procedure as for B·tube.
2)Depression prism unit
For disassembly, refer to Fig. 11 .