Nikon 50i and E100 manual UW·tube 4. ER·tube

Page 93

R-4106.A

DISASSEMBLY PROCEDURE

3. UW·tube 4. ER·tube

 

 

3.UW·tube

Release screw ~ xJ, and separate the binocular body from the depression pnsm unit.

1)Photo·tube

For disassembly, refer to Fig. 7, taking the cautions as below:

When taking off Photo·tube sleeve e±) out of interna! tube G), mark off a line on the sleeve ©to regain the original position.

2)To disassemble the binocular body, refer to Fig. 9 and 1O.

3)To disassemble the depression prism unit, refer to Fig. 8, taking the cautions as below:

Do not remove s1multaneously both the dovetail grooves (!I) from bottom plate @, because either is to be used as fiducial, when reassembling.

4.ER·tube

Release screws @ x J and separa te the binocular body from the depression prism unit.

1)Binocular body

For disassembly, follow the sarne procedure as for B·tube.

2)Depression prism unit

For disassembly, refer to Fig. 11 .

- 23 -

Image 93
Contents ·~.·i~~· OPTIPHOT, Labophot Booy Contents Care for Starting Repair Care for Starting Repai R 4104.ATools ANO Preparations 4104.ASpecial Tools Tools ANO PreparationsSpecial Tools Special Toais Optiphot X. XP Lubricant and AdhesiveLubricant and Adhesive ~11~221 ~,f-íli\ ~!X4 ti Tools ANO Prepaaations ~,,.........~2 J21\ Lub~ OPTIPHOT, Labophot Optiphot 41 .A UJ ~\JfsPl Others to be Repaired Mechanism ANO Principle @ .@. · @,@, @, @,@Mechanism ANO Principle Disassembly Procedure Coaxial Coarse and Fine Focus UnitCoaxial Coam and Fine Focus Unit OPTIPHOT, Labophot Fig Coaxial Coami and Fine Focus Unit Disassembl Y ProcedureSubrtage Unit OPTIPHOT, Labophot Subrtage UnitSubstage Unit ~/~Ir ---------~32 5 \ Arm Unit OPTIPHOT, Labophot Arm UnitLubricant G605 H40 Base Unit OPTIPHOT, Labophot Base UnitBase Unit @.Fig Reassembl Y Procedure Reassembly ProcedureCoaxial Coarse and Fine Focus Knob Unit ~~~ 123f109f 105 i Tension gauge 4104 a Reassembly ProceoureReassembl V Procedure Reassembly Procedure Adjustment Torque of Coarse Focus KnobSmooth Turning of the Laft Fine Focus Knob AD.JUSTMENT Smooth Turning of the Right Fine Focus KnobCentering the Tube Attaching and Stage Attaching Parts 4104ACantering of Optical System in the Microscope Base Centering of Optical System in the Microscope Base Tube Length 04.A Adjustment of Brightness Control CircuitAojustment Adjustment of Brightness Contrai Circuitªf -~...~ ~ I,~ Adjustment of Brightness Contrai CircuitTools for General Use Toais for General Use 2. Exclusive ToaisExclusive Tools 4108.ALubricant 4. Adhesive 5. Others AdhesiveOthers Disassembling Procedure Disassembling Paoceoure Disassembling Procedure Disassembling Procedure 4108.A Reassembling the Slide Holder Reassembling ProcedureReassemlbing Procedure Reassembling Procedure Reassembling Procedure Procedure j Repair and AdjustmentFailure of the Slide Glass Retainer Slack of the Slide Gim CarrierRepair and Adjustment ProcedureWobbling in X right and left Movement Procadure Repair ANO AdjustmentRepair ANO Adjustment ProcedureRapar and Adjustment Wobbling in Y to and fro MovementRepair ANO· Adjustment Failure of the X right and left Handle Function Failure of the Y to and fro Handle Function SIKk of the Handle Assembly~.~f ····· ···· Page Care FOR. Starting Repair Care for Starting Repair 4106.AToais for General Use Special Toais 1010Lubricants Tools ANO·PREPARATIONSNo.350 Adhesive 5. Othen to be Repaired 4106 aTube Depression prism ·tube Binocular body ·tubeBinocular body ·~0 Adhcsive ~ll-8 UW·tubePhoto tube UW-tube Oepression prism UW-tube Binocular body 4106 a Tools ANO. p ReparationsUW·tubeBemocular· body ER-tube Oepression prism Tube ·tube UW-tube UW·tubeUW·tube ER·tube Erect image binocular-tuba ER-tube Erect image binooular-tube·tube ·tube 2. F·tubeTube Disassembl Y ProceoureER·tube UW·tube 4. ER·tube~x3 ·tube \~ j Coincide the prism comer w1th thu edgcs @x2 REASSEMBL. Y Procedure·tube UW-tube Th1s edges UW·tuoo ER·tube Reassembl Y ProcedureReference Table for Repair ANO Aojustment Brightneu Fall-offin the Circumferenca of Viewfield Repair ANO Adjustment B, F, ER and UW-tubeProcadure Not Smooth Operation of lnterpupillary Distance AdjustmentRepair ANO Displacement of the Cantar of Right or Left Viewfield ProceduraLevei Differance of Paired Exit Pupils Repair ANO Adjustment Ln the Case of a Larger Displacement of the Cent9r Depression Prism Unit Depnmion Prism Unit Deprmion Prism Unit Binocular Body \\\Coincidence 4106.A Binocular Body