R.4104.A

REASSEMBLY PROCEDURE

2. Substage Unit

 

 

2.Substage Unit

For reassembly, reverse the disassembly procedure, taking the cautions as below:

1)The positions to be oiled or cemented should be cleaned and dried, beforehand. Use only the specified lubricant :md adhesive.

2)Rotation torque of the condenser focus knob @ is to be adjusted to 450g - 500g, when the knob is turned with pinion @ held by the hand. \1ake the adjustment by

turning the pinion bearing @ . (Measuring procedure: Fig. 38)

After adjustment, fix the knob with set screw @ .

3)If in Disassembly Process (D (P. 24) both the dovetails have been removed simultaneousty for an unavoidable reason, it will be necessary for finding the fiducial position to use the specified positioning too!.

4)Check for smooth movement of the sliding parts of the dovetails, without slack, unevenness or the like.

Tension gauge

Fig. 38

5)Proceed to fitting of the rack as below:

i)Insert set screws @x2 into the main body of substage.

ii)Attach the knob to @ with @x 2, transitorily.

üi) Attach rack@ to dovetail (male)@ with @x2.

iv)After releasing @ x 2 and inserting the dovetail, tum pinion bearing @ , so that lhe gearing of rack @ is achieved, and fix it in this position with @ x 2 (utilizíng lhe eccentricity of @ ).

v)Make sure of no clearance between @ and @ and smooth operation without slack, unevenness or the like.

- 34 -

Image 37
Contents ·~.·i~~· OPTIPHOT, Labophot Booy Contents Care for Starting Repair Care for Starting Repai R 4104.ATools ANO Preparations 4104.ASpecial Tools Tools ANO PreparationsSpecial Tools Special Toais Optiphot X. XP Lubricant and AdhesiveLubricant and Adhesive ~11~221 ~,f-íli\ ~!X4 ti Tools ANO Prepaaations ~,,.........~2 J21\ Lub~ OPTIPHOT, Labophot Optiphot 41 .A UJ ~\JfsPl Others to be Repaired Mechanism ANO Principle @ .@. · @,@, @, @,@Mechanism ANO Principle Coaxial Coarse and Fine Focus Unit Disassembly ProcedureCoaxial Coam and Fine Focus Unit OPTIPHOT, Labophot Fig Coaxial Coami and Fine Focus Unit Disassembl Y ProcedureSubrtage Unit OPTIPHOT, Labophot Subrtage UnitSubstage Unit ~/~Ir ---------~32 5 \ Arm Unit OPTIPHOT, Labophot Arm UnitLubricant G605 H40 Base Unit OPTIPHOT, Labophot Base UnitBase Unit @.Fig Reassembly Procedure Reassembl Y ProcedureCoaxial Coarse and Fine Focus Knob Unit ~~~ 123f109f 105 i Tension gauge 4104 a Reassembly ProceoureReassembl V Procedure Reassembly Procedure Torque of Coarse Focus Knob AdjustmentSmooth Turning of the Laft Fine Focus Knob AD.JUSTMENT Smooth Turning of the Right Fine Focus KnobCentering the Tube Attaching and Stage Attaching Parts 4104ACantering of Optical System in the Microscope Base Centering of Optical System in the Microscope Base Tube Length 04.A Adjustment of Brightness Control CircuitAojustment Adjustment of Brightness Contrai Circuitªf -~...~ ~ I,~ Adjustment of Brightness Contrai CircuitTools for General Use Toais for General Use 2. Exclusive ToaisExclusive Tools 4108.AAdhesive Lubricant 4. Adhesive 5. OthersOthers Disassembling Procedure Disassembling Paoceoure Disassembling Procedure Disassembling Procedure 4108.A Reassembling the Slide Holder Reassembling ProcedureReassemlbing Procedure Reassembling Procedure Reassembling Procedure Procedure j Repair and AdjustmentFailure of the Slide Glass Retainer Slack of the Slide Gim CarrierProcedure Repair and AdjustmentWobbling in X right and left Movement Procadure Repair ANO AdjustmentRepair ANO Adjustment ProcedureRapar and Adjustment Wobbling in Y to and fro MovementRepair ANO· Adjustment Failure of the X right and left Handle Function Failure of the Y to and fro Handle Function SIKk of the Handle Assembly~.~f ····· ···· Page Care FOR. Starting Repair Care for Starting Repair 4106.AToais for General Use Special Toais 1010Lubricants Tools ANO·PREPARATIONSNo.350 Adhesive 5. Othen to be Repaired 4106 aTube Depression prism ·tube Binocular body ·tubeBinocular body ·~0 Adhcsive ~ll-8 UW·tubePhoto tube UW-tube Oepression prism UW-tube Binocular body Reparations 4106 a Tools ANO. pUW·tubeBemocular· body ER-tube Oepression prism Tube ·tube UW-tube UW·tubeUW·tube ER·tube Erect image binocular-tuba ER-tube Erect image binooular-tube·tube ·tube 2. F·tubeTube Disassembl Y ProceoureER·tube UW·tube 4. ER·tube~x3 ·tube \~ j Coincide the prism comer w1th thu edgcs @x2 REASSEMBL. Y Procedure·tube UW-tube Th1s edges UW·tuoo ER·tube Reassembl Y ProcedureReference Table for Repair ANO Aojustment Brightneu Fall-offin the Circumferenca of Viewfield Repair ANO Adjustment B, F, ER and UW-tubeProcadure Not Smooth Operation of lnterpupillary Distance AdjustmentRepair ANO Procedura Displacement of the Cantar of Right or Left ViewfieldLevei Differance of Paired Exit Pupils Repair ANO Adjustment Ln the Case of a Larger Displacement of the Cent9r Depression Prism Unit Depnmion Prism Unit Deprmion Prism Unit Binocular Body \\\Coincidence 4106.A Binocular Body