Nikon 50i and E100 manual @x2, REASSEMBL. Y Procedure

Page 98

R-4106.A

REASSEMBL. Y PROCEDURE

2. F·tube

 

 

2. F·tube

2-1. Depression prism unit (Refer to Fig. 5 and 6)

Up j_

l)Attach lens charnber G) (with @ and @ inside, to be positioned as ~Tthicker)

to prism holder@. using washer @ and screw @ x 2.

2)lnsert tenses @ and @ (positioned as fb.±. thicker)

o~T

into prism holder ~ and nx them with retaining ring @.

3)Place cylindrical rollers @ x4 and parallel pin 56C on the

fiducial roiler race @) attached to binocular body attaching plateeij) through @x2 (Fig. 23). Apply oil

G7920 to @ and ~

4)Place the lower side of roller race (V onto cylindrical rollers @ of the above cited fiducial

Down

Fig. 22

roilers race @, and @ x 4

 

and @ onto the upper sur·

 

face of roller race @. through

 

@x2.

 

5) Apply adhesive No. 616 to the

~--

contact surface @ (at three

~

points), and attach 9 to

57' ~~

@ withscrews ~x2

A

 

transitorily.

 

 

6)Make positional adjustment of

(V on the upper surface of

Fig. 23

e@, so that@ operates smoothly witltout any slack in respect to roller race @i.2 of

~Fix it with screw ~x 2.

Operating torque of@ is to be SOg - 60g, measured with a tension gauge.

-28-

Image 98
Contents ·~.·i~~· OPTIPHOT, Labophot Booy Contents 4104.A Care for Starting Repair Care for Starting Repai R4104.A Tools ANO PreparationsTools ANO Preparations Special ToolsSpecial Tools Special Toais Lubricant and Adhesive Optiphot X. XP~11 Lubricant and Adhesive~221 ~,f-íli\ ~!X4 ti Tools ANO Prepaaations ~,,.........~2 J21\ Lub~ OPTIPHOT, Labophot Optiphot UJ ~\ 41 .AJfsPl Others to be Repaired @ .@. · @,@, @, @,@ Mechanism ANO PrincipleMechanism ANO Principle Coaxial Coam and Fine Focus Unit OPTIPHOT, Labophot Fig Disassembly ProcedureCoaxial Coarse and Fine Focus Unit Disassembl Y Procedure Coaxial Coami and Fine Focus UnitSubrtage Unit Subrtage Unit OPTIPHOT, Labophot~/~ Substage UnitIr ---------~32 5 \ Arm Unit Arm Unit OPTIPHOT, LabophotLubricant G605 H40 Base Unit Base Unit OPTIPHOT, LabophotBase Unit @.Fig Coaxial Coarse and Fine Focus Knob Unit Reassembl Y ProcedureReassembly Procedure ~~~ 123f109f 105 i Tension gauge Reassembly Proceoure 4104 aReassembl V Procedure Reassembly Procedure Smooth Turning of the Laft Fine Focus Knob AdjustmentTorque of Coarse Focus Knob Smooth Turning of the Right Fine Focus Knob AD.JUSTMENT4104A Centering the Tube Attaching and Stage Attaching PartsCantering of Optical System in the Microscope Base Centering of Optical System in the Microscope Base Tube Length Adjustment of Brightness Control Circuit 04.AAdjustment of Brightness Contrai Circuit Aojustmentªf -~...~ Adjustment of Brightness Contrai Circuit ~ I,~Exclusive Tools Toais for General Use 2. Exclusive ToaisTools for General Use 4108.AOthers Lubricant 4. Adhesive 5. OthersAdhesive Disassembling Procedure Disassembling Paoceoure Disassembling Procedure Disassembling Procedure 4108.A Reassembling Procedure Reassembling the Slide HolderReassemlbing Procedure Reassembling Procedure Reassembling Procedure Failure of the Slide Glass Retainer Repair and AdjustmentProcedure j Slack of the Slide Gim CarrierWobbling in X right and left Movement Repair and AdjustmentProcedure Repair ANO Adjustment ProcadureRapar and Adjustment ProcedureRepair ANO Adjustment Wobbling in Y to and fro MovementRepair ANO· Adjustment Failure of the X right and left Handle Function SIKk of the Handle Assembly Failure of the Y to and fro Handle Function~.~f ····· ···· Page 4106.A Care FOR. Starting Repair Care for Starting Repair1010 Toais for General Use Special ToaisTools ANO·PREPARATIONS LubricantsNo.350 4106 a Adhesive 5. Othen to be RepairedTube Depression prism ·tube Binocular body ·tubeBinocular body ·~0 Adhcsive ~ll-8 UW·tubePhoto tube UW-tube Oepression prism UW-tube Binocular body UW·tubeBemocular· body 4106 a Tools ANO. pReparations ER-tube Oepression prism Tube ·tube UW·tube UW-tubeUW·tube ER-tube Erect image binooular-tube ER·tube Erect image binocular-tubaTube ·tube 2. F·tube·tube Disassembl Y ProceoureUW·tube 4. ER·tube ER·tube~x3 ·tube \~ j Coincide the prism comer w1th thu edgcs REASSEMBL. Y Procedure @x2·tube UW-tube Th1s edges UW·tuoo Reassembl Y Procedure ER·tubeReference Table for Repair ANO Aojustment Repair ANO Adjustment B, F, ER and UW-tube Brightneu Fall-offin the Circumferenca of ViewfieldNot Smooth Operation of lnterpupillary Distance Adjustment ProcadureRepair ANO Levei Differance of Paired Exit Pupils Displacement of the Cantar of Right or Left ViewfieldProcedura Repair ANO Adjustment Ln the Case of a Larger Displacement of the Cent9r Depression Prism Unit Depnmion Prism Unit Deprmion Prism Unit \\\ Binocular BodyCoincidence 4106.A Binocular Body