F-11

TROUBLESHOOTING & REPAIR

F-11

 

 

 

 

 

Observe all Safety Guidelines detailed throughout this manual

 

 

 

 

 

 

 

 

 

PROBLEMS

 

 

POSSIBLE AREAS OF

 

RECOMMENDED

 

(SYMPTOMS)

 

MISADJUSTMENT(S)

 

COURSE OF ACTION

 

 

 

 

 

 

 

 

 

 

 

 

WELDING PROBLEMS

 

 

 

 

 

 

 

 

 

 

Poor arc starting when the DC-

1.

Make sure the proper welding

1.

Check the 3CR reed switch.

 

1000 is in the CV Sub-Arc or CV

 

procedures are being used

 

The voltage from lead #215 to

 

Innershield Modes.

 

 

(wire feed speed, arc voltage

 

lead #220 should be 8 VDC

 

 

 

 

and wire size).

 

when the DC-1000 is in an

 

 

 

 

 

 

idle condition (on but not

 

 

 

2.

Check weld cables for loose

 

welding). When the machine

 

 

 

 

or faulty connections.

 

is producing welding current,

 

 

 

 

 

 

the reed switch (3CR) should

 

 

 

 

 

 

close and the voltage from

 

 

 

 

 

 

#215 to #220 should drop to

 

 

 

 

 

 

zero.

 

 

 

 

 

 

2.

Perform the Firing Board

 

 

 

 

 

 

Test.

 

 

 

 

 

 

3

Perform the

SCR Output

 

 

 

 

 

 

Bridge Test.

 

 

 

 

 

 

4.

The control board may be

 

 

 

 

 

 

faulty - Replace.

 

 

 

 

 

 

 

 

Poor arc characteristics in all

1.

Check for the correct input

1.

Check the

Mode Switch

 

processes.

 

 

voltages on the three phase

 

(SW4) and the associated

 

 

 

 

input lines at the DC-1000.

 

wiring for loose or faulty con-

 

 

 

 

 

 

nections. See wiring diagram.

 

 

 

2. Make sure the proper welding

 

 

 

 

 

 

 

procedures are being used

2.

Check the 3CR reed switch.

 

 

 

 

(wire feed speed, arc voltage

 

The voltage from lead #215 to

 

 

 

 

and wire size).

 

lead #220 should be 8 VDC

 

 

 

 

 

 

when the DC-1000 is in an

 

 

 

3.

Check the welding cables for

 

idle condition. When the

 

 

 

 

loose or faulty connections.

 

machine is producing welding

 

 

 

 

 

 

current, the reed switch (3CR)

 

 

 

 

 

 

should close and the voltage

 

 

 

 

 

 

from #215 to #220 should

 

 

 

 

 

 

drop to zero.

 

 

 

 

 

 

3.

Perform the Firing Board

 

 

 

 

 

 

Test.

 

 

 

 

 

 

4.

Perform the

SCR Output

 

 

 

 

 

 

Bridge Test.

 

 

 

 

 

 

5.

The control board may be

 

 

 

 

 

 

faulty - Replace.

CAUTION

If for any reason you do not understand the test procedures or are unable to perform the tests/repairs safely, contact the Lincoln Electric Service Department for technical troubleshooting assistance before you proceed. Call 216-383-2531 or 1-800-833-9353.

IDEALARC DC-1000

Image 51
Contents Idealarc Electric and Magnetic Fields SafetyElectric Shock can kill Welding Sparks can cause fire or explosion Sûreté Pour Soudage a L’Arc Précautions DE SûretéMaster Table of Contents for ALL Sections Installation Output Technical Specifications Idealarc DCStacking Electric Shock can KillSelect Proper Location Electrical Input ConnectionsGround Connection Reconnect ProcedureFuse and Wire Sizes Input Power Supply ConnectionsFigure A.3a Input Connection Diagram Figure A.3c -Input Connection Diagram Figure A.3d Input Supply Connection Diagram Output Connections Cutting Operation Connection for AIR/CARBON ARCIdealarc DC-1000 Operation Electric Shock can kill Safety InstructionsRecommended Processes and Equipment Operational Features and ControlsDesign Features Advantages General DescriptionWelding Capability Return to Section Controls and SettingsWelding Procedure Recommendations Remote Control of Machine OperationNA-3 Automatic Wire Feeder ARC Striking with Idealarc DC-1000 and the NA-3 Start Board Remote LN-8 SEMI-AUTOMATIC Wire FeederTable of Contents AccessoriesRemote Output Control Optional OPTIONS/ACCESSORIESConnecting the NA-3 or LT-7 to Idealarc DC-1000Figure C.2 NA-5 Wire Feeder Connection to Idealarc DC-1000 Connecting the NA-5 to the Idealarc DC-1000Connecting the LN-8 to the Idealarc DC-1000 Connecting the LN-9 Wire Feeder to the Idealarc DC-1000 Connecting the LN-7 Wire Feeder to the Idealarc DC-1000 Maintenance Routine and Periodic Maintenance Safety Precautions Electric Shock Can KillFigure D.1 Major Component Locations Idealarc DC-1000 Theory of Operation TOR, and Main Transformer Input Line VOLTAGE, ContacCONTROL, and Feedback Output RectificationCircuits Contactor HOLD-IN Protection DevicesFigure E.5 SCR Operation SCR OperationIdealarc DC-1000 Troubleshooting & Repair Locate Problem Symptom HOW to USE Troubleshooting GuideBoard can be damaged by static electricity PC Board Troubleshooting ProceduresControl Transformer Test Return Return to Master TOC Return Return to Section TOC Perform SCR Output Bridge Troubleshooting & Repair Function Problems Perform Firing Board Test Perform Control Board Test SCR Output Troubleshooting & Repair Control Transformer T2 Voltage Test Test DescriptionFigure F.1 Control Transformer Lead Location Test ProcedureFigure F.2 Control Transformer X1 and X2 Test Points Main Transformer T1 Voltage Test Test Description Figure F.4 Main Secondary Lead Test Points From Expected VAC Firing Board Test Test Procedure for Normal Firing Board Operation Figure F.7 Terminal Strip Jumper Wire Connections Figure F.8 Output Pilot Relay 4CR Location Then Table F.1 LED 7, 8 and 9 Check ListControl Board Test Figure F.9 Control Board LED and Pin Locations Test Procedure for Normal Control Board OperationIf LED 1 does not light, when the start switch is on Possible Problems Pertaining to the Control BoardFigure F.11 LED 2 Open Circuit Voltage Test Points If LED 4 lights Static SCR Test Figure F.12 Test ProcedureFigure F.13 SCR Heat Sink Assembly Test Points Static SCR Test Active SCR Test Active SCR Test Figure F.16 Heat Sink Assembly Test Points Battery Test Scope Settings Troubleshooting & Repair Troubleshooting & Repair Troubleshooting & Repair Troubleshooting & Repair Troubleshooting & Repair Troubleshooting & Repair Input Contactor 1CR Cleaning AND/OR Replacement Figure F.18 Input Contactor Cover Removal Cleaning ProcedureRemoval Contactor ReplacementInstallation SCR Output Bridge Replacement Figure F.22 Preparation For SCR Output Bridge Removal SCR Output Bridge RemovalFigure F.23 SCR Output Bridge Removal Removal of Individual SCR Heat Sink Assemblies SCR Output Bridge Installation SCR Heat Sink InstallationTOC Removal and Reassembly of Lift Bail Figure F.25 Lift Bail Removal Removal of Lift BailReassembly of Lift Bail Main Transformer Removal and Installation Figure F.26 Main Transformer Disassembly Main Transformer Disassembly and AssemblyCoil Removal Replacement Main Transformer DisassemblyTransformer Reassembly Transformer InstallationIdealarc DC-1000 Open Circuit Voltages Input Idle Amps and WattsIdealarc DC-1000 Electrical Diagrams Idealarc DC-1000 Wiring Diagram Codes 9919 Operating Schematic Control PC Board G1585 Layout Firing PC Board G1486-5 and above Layout Control PC Board G1585 Schematic Firing PC Board G1486-5 and above Schematic Output Snubber M14312 Schematic