Diagnostics and Service
Table 24: Humidity versus voltage.
RH (%) | VDC (mV) | RH (%) | VDC (mV) |
10 | 1330 | 55 | 2480 |
15 | 1475 | 60 | 2600 |
20 | 1610 | 65 | 2730 |
25 | 1740 | 70 | 2860 |
30 | 1870 | 75 | 2980 |
35 | 1995 | 80 | 3115 |
40 | 2120 | 85 | 3250 |
45 | 2235 | 90 | 3390 |
50 | 2360 | 95 | 3530 |
Troubleshooting Carbon Dioxide (CO2) Sensors
The UVC is configured to use a
Use the following procedure to troubleshoot a suspect sensor.
1Disconnect the sensors output voltage lead from the UVC analog input
2Using some other calibrated CO2 sensing device, take a CO2 reading at the sensor location.
3Use the CO2 reading from Step 2 to determine the expected sensor voltage from Table 25.
4Using a calibrated
5Compare the expected voltage to the actual voltage.
6If the actual voltage value deviates substantially (more than 10%) from the expected voltage, replace the sensor.
In the unlikely event that the CO2 sensor requires calibration, consult the factory for information on obtaining calibration equipment and instructions.
Table 25: CO2 versus voltage table
CO2 (PPM) | VDC (V) | CO2 (PPM) | VDC (V) |
300 | 1.5 | 1200 | 6.0 |
400 | 2.0 | 1300 | 6.5 |
500 | 2.5 | 1400 | 7.0 |
600 | 3.0 | 1500 | 7.5 |
700 | 3.5 | 1600 | 8.0 |
800 | 4.0 | 1700 | 8.5 |
900 | 4.5 | 1800 | 9.0 |
1000 | 5.0 | 1900 | 9.5 |
46 | McQuay OM 751 |