Omega Engineering FP-6000 manual Cap nuts Installation tool

Page 21

installation tool threaded shaft

cap nuts

installation tool

protector plate

bearing plate

hex nuts

 

sensor cable

swivel mount

 

w/cable port

upper hex nuts

sensor flange

 

1 lower hex nut and jam nut

sensor body

5.Wearing safety face protection, turn the installation tool shaft counterclockwise to withdraw sensor until the sensor flange contacts the upper hex nuts.

6.Raise one lower hex and jam nut to bottom of sensor flange.

7.Close valve, remove bearing plate and tool.

To remove the sensor

8.Wearing safety face protection, cover the bleed valve with suitable protection (rag, towel, etc.) and open the bleed valve (ccw rotation) to relieve internal pressure. Pull sensor up until bleed valve purges some fluid (indicating sensor is past 1st o-ring seal inside brass sensor nut) then remove sensor from brass sensor nut/threaded rod assembly.

Caution: In case of a leaky isolation valve, the sensor will be under a slight amount of pressure. Care should be taken when removing the sensor. Use the bleed valve to relieve this pressure taking care not to spray fluid on yourself or others.

When reinstalling the sensor: leave one lower hex nut in position to guide sensor to proper height before opening valve. Return to "H" dimension height after valve is opened.

15

Image 21
Contents User’s Guide Servicing North America For immediate technical or application assistanceServicing Europe Table of Contents Page Important Safety Information Type Omega FP-6000 Series Warranty RecordChapter Introduction DescriptionTheory of Operation Chapter Installation and Wiring Location of FittingSensor Position +30 Sensor WiringHardware, Standard Sensor Standard Fitting InstallationHardware, Hot-Tap Sensor 11/4 in mm diameter drill Pipe thread sealant Tape measureHot-Tap Fitting Installation Remove brass sensor nut from sensorThread brass sensor nut into pipe fitting Wall Thickness Calculating the H DimensionHot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.D Record your pipes H dimension for future referenceStandard Installation Sensor flange Alignment rod Direction of flow Hot-Tap Installation StepThread one jam nut and lower hex nut onto Direction of flow Make sure the bleed valve is closed full clockwise positionUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Standard Sensor Removal ChapterHot-Tap Sensor Removal Into the port to prevent damageCap nuts Installation tool Chapter Maintenance and Replacement Parts MaintenanceReplacement Parts Rotor Replacement Procedure Dimensions for Standard Sensors DimensionsWrought Steel Pipe Per Ansi Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inchesDimensions for Hot-Tap Sensors GPM/HZ LPM/HZ Factors Stainless SteelFactors Wrought Steel GPM/HZ LPM/HZ Factors Plastic PipeGeneral Data SpecificationsSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course