Chapter 4

Sensor Removal Procedures

4.1 Standard Sensor Removal

To remove the Standard FP-6000 Series sensor from a depressurized empty pipe, simply remove the cap nuts and upper hex nuts located above the sensor flange. Pull up on sensor flange with twisting motion.

4.2 Hot-Tap Sensor Removal

protector plate cap nuts

protector plate

protector plate hex nut

To remove the Hot-Tap sensor safely from a pressurized active pipe, the entire installation process must be reversed.

1.Remove the cap nuts, protector plate, protector plate hex nuts, and sensor cable grommet.

2.Thread installation tool in place and secure bearing plate in place of sensor protector plate.

3.Turn shaft of installation tool clockwise to lower tool into opening in sensor flange. Guide cable

into the port to prevent damage.

installation tool threaded shaft

upper hex nuts

and lock washers

sensor flange

14.2 in.

lower hex and

(361 mm)

jam nuts

4. Wearing

safety face protection,

loosen the upper

hex nuts and raise to 14.2 in.

(361 mm) from top of brass sensor nut to bottom of nut. This measurement is critical to maintain watertight seal in sensor while allowing clearance to close the isolation valve.

process pipe (side view)

14

Image 20
Contents User’s Guide Servicing Europe Servicing North AmericaFor immediate technical or application assistance Table of Contents Page Important Safety Information Warranty Record Type Omega FP-6000 SeriesTheory of Operation Chapter IntroductionDescription Sensor Position Chapter Installation and WiringLocation of Fitting Sensor Wiring +30Standard Fitting Installation Hardware, Standard SensorHardware, Hot-Tap Sensor 11/4 in mm diameter drill Pipe thread sealant Tape measureThread brass sensor nut into pipe fitting Hot-Tap Fitting InstallationRemove brass sensor nut from sensor Calculating the H Dimension Wall ThicknessRecord your pipes H dimension for future reference Hot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.DStandard Installation Sensor flange Alignment rod Direction of flow Thread one jam nut and lower hex nut onto Hot-Tap InstallationStep Make sure the bleed valve is closed full clockwise position Direction of flowUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Chapter Standard Sensor RemovalHot-Tap Sensor Removal Into the port to prevent damageCap nuts Installation tool Replacement Parts Chapter Maintenance and Replacement PartsMaintenance Rotor Replacement Procedure Dimensions Dimensions for Standard SensorsWrought Steel Pipe Per Ansi Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inchesDimensions for Hot-Tap Sensors Factors Stainless Steel GPM/HZ LPM/HZFactors Wrought Steel Factors Plastic Pipe GPM/HZ LPM/HZSpecifications General DataSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course