3.Align the sensor cable with the swivel mount cable port to prevent cable pinching. Use a 3/8 inch wrench or socket to turn the installation tool shaft clockwise until it is seated in the hole at the top of the sensor flange.

4.Wearing safety face protection, slowly open the isolation valve to the full open position. Loosen the lower hex and jam nuts and move them to the required "H" dimension. Turn the installation tool shaft clockwise until the sensor flange contacts the lower hex and jam nuts. Thread the upper hex nuts down until they contact the sensor flange. Tighten the upper hex nuts to secure the sensor.

5.Remove cap nuts and withdraw the installation tool by turning shaft counterclockwise. Be careful to not damage cable. Snap cable grommet into top of sensor and replace protector plate and cap nuts.

protector plate

installation tool shaft

upper hex nuts

cap nuts

alignment rod

lower hex nuts jam nuts

cap nuts

protector plate

protector plate hex nut

"H"

direction of flow

13

Image 19 Contents
User’s Guide For immediate technical or application assistance Servicing North AmericaServicing Europe Table of Contents Page Important Safety Information Type Omega FP-6000 Series Warranty RecordDescription Chapter IntroductionTheory of Operation Location of Fitting Chapter Installation and WiringSensor Position +30 Sensor Wiring11/4 in mm diameter drill Pipe thread sealant Tape measure Standard Fitting InstallationHardware, Standard Sensor Hardware, Hot-Tap SensorRemove brass sensor nut from sensor Hot-Tap Fitting InstallationThread brass sensor nut into pipe fitting Wall Thickness Calculating the H DimensionHot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.D Record your pipes H dimension for future referenceStandard Installation Sensor flange Alignment rod Direction of flow Step Hot-Tap InstallationThread one jam nut and lower hex nut onto Direction of flow Make sure the bleed valve is closed full clockwise positionUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Into the port to prevent damage ChapterStandard Sensor Removal Hot-Tap Sensor RemovalCap nuts Installation tool Maintenance Chapter Maintenance and Replacement PartsReplacement Parts Rotor Replacement Procedure Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inches DimensionsDimensions for Standard Sensors Wrought Steel Pipe Per AnsiDimensions for Hot-Tap Sensors GPM/HZ LPM/HZ Factors Stainless SteelFactors Wrought Steel GPM/HZ LPM/HZ Factors Plastic PipeGeneral Data SpecificationsSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course
Related manuals
Manual 32 pages 56.58 Kb