Omega Engineering FP-6000 manual Important Safety Information

Page 5

Important Safety Information!

CAUTION:

(Standard version) Never remove the flow sensor from a pressurized pipe. Always wear safety face protection during sensor installation/removal.

(Hot-Tap version) Follow the recommended installation/removal instructions in this manual. Always wear safety face protection during sensor installation/removal.

Pipe fittings MUST be installed by a certified welder only. OMEGA will not assume liability of any kind for improper fitting installations.

Serious bodily injury and/or damage to the sensor can result if the conditions and specifications presented in this manual are exceeded. DO NOT exceed specifications under any circumstances.

The FP-6000 Hot-Tap system's overall specifica- tions and limitations depend on the lowest maxi- mum rating of the components associated with the system. In other words, the Hot-Tap system is only as strong as its weakest link. For example, a ball valve, a component of the system, is rated at a maximum 100 psi @ 185 °F, limiting the entire system's maximum pressure/temperature rating to 100 psi @ 185 °F. All higher maximum specifica- tions MUST yield to the component with the lowest maximum specification.

Maximum Operating Pressure: 225 psi (15 bar)

Maximum Operating Temperature:

212 °F (100 °C)

Image 5 Contents
User’s Guide Servicing Europe Servicing North AmericaFor immediate technical or application assistance Table of Contents Page Important Safety Information Type Omega FP-6000 Series Warranty RecordTheory of Operation Chapter IntroductionDescription Sensor Position Chapter Installation and WiringLocation of Fitting +30 Sensor WiringHardware, Standard Sensor Standard Fitting InstallationHardware, Hot-Tap Sensor 11/4 in mm diameter drill Pipe thread sealant Tape measureThread brass sensor nut into pipe fitting Hot-Tap Fitting InstallationRemove brass sensor nut from sensor Wall Thickness Calculating the H DimensionHot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.D Record your pipes H dimension for future referenceStandard Installation Sensor flange Alignment rod Direction of flow Thread one jam nut and lower hex nut onto Hot-Tap InstallationStep Direction of flow Make sure the bleed valve is closed full clockwise positionUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Standard Sensor Removal ChapterHot-Tap Sensor Removal Into the port to prevent damageCap nuts Installation tool Replacement Parts Chapter Maintenance and Replacement PartsMaintenance Rotor Replacement Procedure Dimensions for Standard Sensors DimensionsWrought Steel Pipe Per Ansi Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inchesDimensions for Hot-Tap Sensors GPM/HZ LPM/HZ Factors Stainless SteelFactors Wrought Steel GPM/HZ LPM/HZ Factors Plastic PipeGeneral Data SpecificationsSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course