Omega Engineering FP-6000 manual Record your pipes H dimension for future reference

Page 13

Standard Sensor

H = 5.95 - pipe wall thickness - (0.10 X I.D.)

Example:

3.0inch schedule 80 wrought steel; Wall thickness= 0.3 in.

Inside diameter = 2.9 in.

H = 5.95 - 0.3 - (0.10 X 2.9)

H = 5.36 in.

sensor flange

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

"H"

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

pipe side view

 

 

 

 

 

 

 

 

 

Record your pipes "H" dimension for future reference:

H= _________________

alignment rod

process pipe

direction of flow

Hot-Tap Sensor

H = 15.00 - pipe wall thickness - (0.10 X I.D.)

protector plate

Example:

10 inch schedule 40 wrought steel; Wall thickness= 0.365 in.

Inside diameter = 10.02 in.

H = 15.00 - 0.365 - (0.10 X 10.02) H = 13.633 in.

Record your pipes "H" dimension for future reference:

"H"

direction of flow

alignment rod

H= _________________

process pipe (side view)

7

Image 13
Contents User’s Guide For immediate technical or application assistance Servicing North AmericaServicing Europe Table of Contents Page Important Safety Information Type Omega FP-6000 Series Warranty RecordDescription Chapter IntroductionTheory of Operation Location of Fitting Chapter Installation and WiringSensor Position +30 Sensor WiringHardware, Standard Sensor Standard Fitting InstallationHardware, Hot-Tap Sensor 11/4 in mm diameter drill Pipe thread sealant Tape measureRemove brass sensor nut from sensor Hot-Tap Fitting InstallationThread brass sensor nut into pipe fitting Wall Thickness Calculating the H DimensionHot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.D Record your pipes H dimension for future referenceStandard Installation Sensor flange Alignment rod Direction of flow Step Hot-Tap InstallationThread one jam nut and lower hex nut onto Direction of flow Make sure the bleed valve is closed full clockwise positionUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Standard Sensor Removal ChapterHot-Tap Sensor Removal Into the port to prevent damageCap nuts Installation tool Maintenance Chapter Maintenance and Replacement PartsReplacement Parts Rotor Replacement Procedure Dimensions for Standard Sensors DimensionsWrought Steel Pipe Per Ansi Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inchesDimensions for Hot-Tap Sensors GPM/HZ LPM/HZ Factors Stainless SteelFactors Wrought Steel GPM/HZ LPM/HZ Factors Plastic PipeGeneral Data SpecificationsSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course