3.7 Hot-Tap Installation

Step 1

hex nut lock washer

brass sensor nut

Step 2

1.Thread one hex nut onto each of the three threaded rods included in package. Install threaded rod with a lock-washer onto the brass sensor nut. Secure rods in place by tightening each hex nut against the brass sensor nut.

2.Thread one jam nut and lower hex nut onto

each stud so the top surface of each nut is 13.75 in. (350 mm) from top surface of brass sensor nut. Secure each hex nut with a jam nut.

This setting is critical to ensure an adequate sensor seal and to prevent the rotor from hitting the isolation valve orifice during installation.

lower hex nut, 2nd jam nut, 1'st

13.75 in.

(350 mm)

lower hex nuts (3/16 x 1/4-20)

jam nuts

(5/32 x 1/4-20)

brass sensor nut

10

Image 16
Contents User’s Guide For immediate technical or application assistance Servicing North AmericaServicing Europe Table of Contents Page Important Safety Information Warranty Record Type Omega FP-6000 SeriesDescription Chapter IntroductionTheory of Operation Location of Fitting Chapter Installation and WiringSensor Position Sensor Wiring +30Standard Fitting Installation Hardware, Standard SensorHardware, Hot-Tap Sensor 11/4 in mm diameter drill Pipe thread sealant Tape measureRemove brass sensor nut from sensor Hot-Tap Fitting InstallationThread brass sensor nut into pipe fitting Calculating the H Dimension Wall ThicknessRecord your pipes H dimension for future reference Hot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.DStandard Installation Sensor flange Alignment rod Direction of flow Step Hot-Tap InstallationThread one jam nut and lower hex nut onto Make sure the bleed valve is closed full clockwise position Direction of flowUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Chapter Standard Sensor RemovalHot-Tap Sensor Removal Into the port to prevent damageCap nuts Installation tool Maintenance Chapter Maintenance and Replacement PartsReplacement Parts Rotor Replacement Procedure Dimensions Dimensions for Standard SensorsWrought Steel Pipe Per Ansi Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inchesDimensions for Hot-Tap Sensors Factors Stainless Steel GPM/HZ LPM/HZFactors Wrought Steel Factors Plastic Pipe GPM/HZ LPM/HZSpecifications General DataSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course