(max. 30°) will help to avoid these problem areas, but use caution. Excessive angles will cause bearing drag at lower flow rates.

On a vertical pipe run locate the sensor where the flow is upward. If downward flow is necessary the system must be designed to prevent air/water vapor pockets from developing in the pipe which will affect the performance of the sensor.

-30°

0°

+30°

Maximum Sensor

Installation Range

Process Pipe

Figure 2

Sensor Installation Range

Vertical mounting is recom- mended to provide best overall performance.

Special Considerations

For Hot-Tap installations allow at least 3 feet of vertical clearance for sensor installation plus the distance required for the isolation valve and fittings attached to the pipe. More clearance may be necessary to suit the drilling machine used during sensor installation.

2.3 Sensor Wiring

Both Standard and Hot-Tap sensor versions include 25 feet of cable. The cable may be extended up to 200 feet without amplification. A 1/2 inch conduit port is available in the sensor to install the cable in protective conduit.

1/2 in. conduit port

black (AC signal out)

red (AC signal out)

silver (shield)

blk

F-

red

F+

shld

instrument

3

Image 9 Contents
User’s Guide Servicing North America For immediate technical or application assistanceServicing Europe Table of Contents Page Important Safety Information Type Omega FP-6000 Series Warranty RecordChapter Introduction DescriptionTheory of Operation Chapter Installation and Wiring Location of FittingSensor Position +30 Sensor WiringHardware, Standard Sensor Standard Fitting InstallationHardware, Hot-Tap Sensor 11/4 in mm diameter drill Pipe thread sealant Tape measureHot-Tap Fitting Installation Remove brass sensor nut from sensorThread brass sensor nut into pipe fitting Wall Thickness Calculating the H DimensionHot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.D Record your pipes H dimension for future referenceStandard Installation Sensor flange Alignment rod Direction of flow Hot-Tap Installation StepThread one jam nut and lower hex nut onto Direction of flow Make sure the bleed valve is closed full clockwise positionUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Standard Sensor Removal ChapterHot-Tap Sensor Removal Into the port to prevent damageCap nuts Installation tool Chapter Maintenance and Replacement Parts MaintenanceReplacement Parts Rotor Replacement Procedure Dimensions for Standard Sensors DimensionsWrought Steel Pipe Per Ansi Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inchesDimensions for Hot-Tap Sensors GPM/HZ LPM/HZ Factors Stainless SteelFactors Wrought Steel GPM/HZ LPM/HZ Factors Plastic PipeGeneral Data SpecificationsSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course