4. Remove brass sensor nut from sensor.

brass sensor nut

pipe fitting

process pipe

(Teflon tape recommended)

5. Thread brass sensor nut into pipe fitting.

3.4 Hot-Tap Fitting Installation

1.Install the pipe fitting on the outside diameter of the pipe according to the manufacturer's instruc- tions. Failure to follow these instructions may result in bodily injury and/or product failure.

2.Install the pipe nipple and isolation valve (ball or gate valve) onto the external pipe fitting using pipe sealant on the threads.

customer supplied ball or gate valve

customer supplied nipple; 1.25 x 2 in. (32 x 50 mm) long

process pipe (side view)

3.Wearing safety face protection, install an appropriate hole cutting tool per manufacturers instructions (e.g. Mueller drilling machine) with a

11/4 in. (32 mm) drill onto the top of the isolation valve, ensuring a tight fit. Use the recommended drill bit size or damage to the isolation valve may occur.

5

Image 11
Contents User’s Guide Servicing Europe Servicing North AmericaFor immediate technical or application assistance Table of Contents Page Important Safety Information Type Omega FP-6000 Series Warranty RecordTheory of Operation Chapter IntroductionDescription Sensor Position Chapter Installation and WiringLocation of Fitting +30 Sensor Wiring11/4 in mm diameter drill Pipe thread sealant Tape measure Standard Fitting InstallationHardware, Standard Sensor Hardware, Hot-Tap SensorThread brass sensor nut into pipe fitting Hot-Tap Fitting InstallationRemove brass sensor nut from sensor Wall Thickness Calculating the H DimensionHot-Tap Sensor = 15.00 pipe wall thickness 0.10 X I.D Record your pipes H dimension for future referenceStandard Installation Sensor flange Alignment rod Direction of flow Thread one jam nut and lower hex nut onto Hot-Tap InstallationStep Direction of flow Make sure the bleed valve is closed full clockwise positionUsing the Hot-Tap Installation Tool Protector plate Installation tool shaft Upper hex nuts Into the port to prevent damage ChapterStandard Sensor Removal Hot-Tap Sensor RemovalCap nuts Installation tool Replacement Parts Chapter Maintenance and Replacement PartsMaintenance Rotor Replacement Procedure Conversion Stainless Steel Pipe Per Ansi B36.19 Mm = inches DimensionsDimensions for Standard Sensors Wrought Steel Pipe Per AnsiDimensions for Hot-Tap Sensors GPM/HZ LPM/HZ Factors Stainless SteelFactors Wrought Steel GPM/HZ LPM/HZ Factors Plastic PipeGeneral Data SpecificationsSpecifications WARRANTY/DISCLAIMER OMEGA…Of Course