Chapter 4 Startup and Shutdown
4.2.3Additional water supply
[TS] switch (pump manual operation switch)
This unit has the [TS] switch (pump manual operation switch) to purge air from the circulating fluid circuit at initial startup. Remove the front panel, and purge air according to the procedures given below.
[TS] switch (green)
Front
Fig. 4-3 [TS] switch (pump manual operation switch)
1. Press the [TS] switch for several seconds, monitoring the fluid level gauge.
Air is purged from the pipe, and the fluid level is lowered.
zThe [ALARM1] lamp comes on (error). The lamp goes out with the press of the [RS] switch after the circulating fluid reaches the range specified on the level indicator.
zSee “Chapter 5 Error Message and Troubleshooting” for error messages.
Control panel
Reset SW [RS] (yellow button)
Front
Fig. 4-4 Location of Reset Switch (Yellow Button)
2. Supply the circulating fluid again according to section “3.2.7Supply of circulating fluid”.
If leakage occurs due to faulty piping including an opened fitting of eternal piping, stop manual operation of the pump and fix the leak.