7. CALIBRATION

7. CALIBRATION

This function adjusts the balance for accurate weighing. Perform a calibration in the following cases.

‰ When the balance is first used.

Calibration (CAL) switch

‰ When the balance has been moved. ‰ For regular calibration.

Calibration (CAL) switch cover 7-1. Calibration using a weight

7-1. Calibration using a weight

‰Prepare the calibration weight (sold separately) in advance.

1. Warm up the balance for at least half an hour with

Press and hold

the CAL switch.

nothing on the pan.

 

 

 

 

 

2. Press and hold the calibration (CAL) switch until

Cal

Cal

appears, and release the switch.

Release the CAL switch.

 

 

 

 

3. The balance displays

Cal 0 .

 

To change the calibration weight value, proceed to

Cal 0

step 4.

 

 

 

To use the calibration weight value in the balance

 

memory, proceed to step 5.

 

4. Press the

SAMPLE

key. The display shows the

060000.

calibration weight value in “gram” that is stored in the

balance. Use the following keys to change the value.

Set the weight using

 

 

 

 

SAMPLE

key

To

select the digit blinking to

the relevant keys.

 

change.

RE-ZERO

key

To set the value of the digit

 

 

 

selected.

 

key

To store the value and return to

PRINT

 

 

 

step 3.

Cal 0

UNITS

key

To cancel the value and return to

 

step 3.

14