9-6. EJ-13 Density Determination Kit

9-6. EJ-13 Density Determination Kit

Using this

option and calculation

Fixing frame

Weighing pan

program, the balance can determine

 

 

the density

(specific gravity) of a

 

 

sample.

 

 

 

EJ-13 Installation

‰Push the fixing frame out, and

remove the weighing pan.

‰Attach the beaker stand to the balance and fit the pan stand to the pan support of the balance.

Sample weighing pan

Upper pan to weigh in air

EJ-13 Installation Pan stand

Lower pan to weigh in water

Pan support

Beaker

‰Place a beaker filled with water on the beaker stand and place the sample weighing pan on top of the pan stand.

Density (specific gravity) measurement

‰The density of a liquid can be changed and there are two ways of setting. One is to set the water temperature and the other is to set density value directly.

‰The factory setting for density of a liquid is 25 °C as water temperature (the density value ρ = 0.99704 (g/cm3) is used to calculate).

‰The density (specific gravity) is calculated by the following formula.

S = A X ρ

A-B

S:Density (specific gravity) of a sample

A:Weight in air

B:Weight in liquid

ρ: Density of liquid (water)

‰The result is shown with two decimal places.