7-2. Gravity acceleration correction

7-2. Gravity acceleration correction

When the balance is first used or has been moved to a different place, it should be calibrated using a calibration weight.

But if the calibration weight cannot be prepared, the gravity acceleration correction will compensate the balance. Change the gravity acceleration value of the balance to the value of the area where the balance will be used. See the gravity acceleration map appended to the end of this manual.

‰Note

Gravity acceleration correction is not required when the balance is calibrated using a calibration weight at the place where the balance is to be used.

1. Press and hold the calibration (CAL) switch until

Press and hold

the CAL switch.

Cal

appears, and release the switch.

 

2.The balance displays Cal 0 .

3.Press the RE-ZERO key.

The display shows the gravity acceleration value stored in the balance.

Use the following keys to change the value.

 

SAMPLE

 

key

To select the digit blinking to

 

 

 

 

 

change.

 

 

 

 

key

To set the value of the digit

 

RE-ZERO

 

 

 

 

 

 

selected.

 

 

key

To store the value and return to

 

PRINT

 

 

 

 

 

step 2.

 

 

key

To cancel the value and return to

 

UNITS

 

 

 

 

 

step 2.

4. After setting the value, press the

 

 

key.

PRINT

 

Cal 0

is displayed again.

 

 

 

5.If it is necessary to calibrate the balance using a calibration weight, go to step 4 of 7-1.

To finish the setting, press UNITS key.

6.end appears and the balance returns to the weighing mode.

Cal

Release the CAL switch.

Cal 0

97985.

Set the value using the relevant keys.

Cal 0To store

Cal 0

Returns to

the weighing mode

16

Page 18