A&D EJ-3000, EJ-6100, EJ-440, EJ-200, EJ-4100 manual Calibration test using a weight, Title, End block

Models: EJ-4100

1 42
Download 42 pages 61.46 Kb
Page 34
Image 34
Title

“Title block” and “End block”

‰When a weight value is recorded as the GLP data, “Title block” and “End block” are added at a group of weight values in the GLP report.

‰To output the GLP report to the printer AD-8121B , use MODE 3 of it.

1. With the weight data displayed, press and hold the PRINT key until start is displayed. Then, the balance outputs the “Title block”.

2.The balance can output the weighing data by pressing the PRINT key or selecting the auto-print mode.

3.Press and hold the PRINT key until recend is displayed. Then, the balance

outputs the “End block” and shows

 

end

.

4. Press the

UNITS

key to return to

the normal

weighing mode.

AD-8121 format “info 1”

Title

General format “info 2”

 

 

 

block

 

 

 

A & D

 

 

 

Manufacturer

 

 

 

 

 

 

 

 

 

MODEL

EJ-300

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Model

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

S/N

 

1234567

 

 

 

 

 

 

 

 

 

 

Serial number

 

 

 

 

 

ID

 

ABCDEF

 

 

 

 

 

 

ID number

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

DATE

 

09/03/22

 

 

 

 

 

 

 

 

Date

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

01:23:45 PM

 

 

 

 

 

 

 

 

 

Start time

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WT

+123.45

g

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WT

+213.43

g

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Weight data

 

 

 

WT

+312.44

g

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

WT

+321.42

g

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

END

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

01:23:56 PM

 

 

 

 

 

 

 

 

End time

 

 

 

 

 

 

 

 

SIGNATURE

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Signature

 

 

 

 

 

 

 

- - - - - - - -

End

block

:Space, ASCII 20h

CR :Carriage return, ASCII 0Dh LF :Line feed, ASCII 04Ah

~~~~~~~~~~~A~&~D<CRLF> MODEL~~~~~EJ-610<CRLF> S/N~~~~~~1234567<CRLF> ID~~~~~~~~ABCDEF<CRLF> DATE<CRLF>

<CRLF>

START<CRLF>

TIME<CRLF>

<CRLF>

<CRLF>

WT~~~~+123.45~~g<CRLF> WT~~~~+213.43~~g<CRLF> WT~~~~+312.44~~g<CRLF> WT~~~~+321.42~~g<CRLF> <CRLF>

END<CRLF>

TIME<CRLF>

<CRLF>

SIGNATURE<CRLF>

<CRLF>

<CRLF>

-~-~-~-~-~-~-~-~<CRLF> <CRLF>

<CRLF>

10-3. Calibration test using a weight

‰Calibration test is to confirm the weighing accuracy using a weight and output the results as GLP report.

‰Set the function setting “info 1” or “info 2” to perform the calibration test.

‰Calibration test does not perform actual calibration but is only to test.

1. Press and hold the calibration (CAL) switch.

CC

appears after

 

Cal

. Release the

switch when

CC

 

is displayed.

 

 

 

Pressing and holding the

 

 

and

 

SAMPLE

PRINT

keys will also display

 

CC

 

.

 

 

 

CC

Release the CAL switch.

32

Page 34
Image 34
A&D EJ-3000, EJ-6100, EJ-440, EJ-4100, EJ-120, EJ-2000, EJ-1500 manual Calibration test using a weight, Title, End block