“Title block” and “End block”
When a weight value is recorded as the GLP data, “Title block” and “End block” are added at a group of weight values in the GLP report.
To output the GLP report to the printer
1. With the weight data displayed, press and hold the PRINT key until start is displayed. Then, the balance outputs the “Title block”.
2.The balance can output the weighing data by pressing the PRINT key or selecting the
3.Press and hold the PRINT key until recend is displayed. Then, the balance
outputs the “End block” and shows |
| end | . | ||
4. Press the | UNITS | key to return to | the normal | weighing mode. | |
Title | General format “info 2” | ||||
|
|
| block |
|
|
| A & D |
|
|
| Manufacturer |
|
| ||||||||||||
|
|
|
|
|
|
| ||||||||||||||
MODEL |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
| Model |
|
|
|
|
|
|
|
| ||||
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
S/N |
| 1234567 |
|
|
|
|
|
|
|
| ||||||||||
|
| Serial number |
|
| ||||||||||||||||
|
|
| ||||||||||||||||||
ID |
| ABCDEF |
|
|
|
|
|
| ID number |
|
|
|
|
|
| |||||
|
|
|
|
|
|
|
|
|
|
|
| |||||||||
DATE |
| 09/03/22 |
|
|
|
|
|
|
|
| Date |
|
|
|
|
|
|
|
| |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
01:23:45 PM |
|
|
|
|
|
|
|
|
| Start time |
|
|
|
|
|
|
| |||
|
|
|
|
|
|
|
|
|
|
| ||||||||||
WT | +123.45 | g |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
WT | +213.43 | g |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
|
|
|
|
|
| Weight data |
|
|
| |||||||||||
WT | +312.44 | g |
|
|
|
|
|
|
|
| ||||||||||
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||
WT | +321.42 | g |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |
END |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| ||||||||||
01:23:56 PM |
|
|
|
|
|
|
|
| End time | |||||||||||
|
|
|
|
|
|
|
| |||||||||||||
SIGNATURE |
|
|
|
|
|
|
|
|
|
|
|
|
|
|
| |||||
|
|
|
|
|
|
| Signature | |||||||||||||
|
|
|
|
|
|
|
- - - - - - - -
End
block
:Space, ASCII 20h
CR :Carriage return, ASCII 0Dh LF :Line feed, ASCII 04Ah
~~~~~~~~~~~A~&~D<CRLF>
<CRLF>
START<CRLF>
TIME<CRLF>
<CRLF>
<CRLF>
WT~~~~+123.45~~g<CRLF> WT~~~~+213.43~~g<CRLF> WT~~~~+312.44~~g<CRLF> WT~~~~+321.42~~g<CRLF> <CRLF>
END<CRLF>
TIME<CRLF>
<CRLF>
SIGNATURE<CRLF>
<CRLF>
<CRLF>
<CRLF>
10-3. Calibration test using a weight
Calibration test is to confirm the weighing accuracy using a weight and output the results as GLP report.
Set the function setting “info 1” or “info 2” to perform the calibration test.
Calibration test does not perform actual calibration but is only to test.
1. Press and hold the calibration (CAL) switch.
CC | appears after |
| Cal | . Release the | ||||||
switch when | CC |
| is displayed. |
|
|
| ||||
Pressing and holding the |
|
| and |
| ||||||
SAMPLE | ||||||||||
keys will also display |
| CC |
| . |
|
|
|
CC
Release the CAL switch.
32