A&D EJ-120, EJ-6100, EJ-440, EJ-3000, EJ-4100, EJ-2000, EJ-1500 manual Cal 50000, 0end000, ‰ Note

Models: EJ-4100

1 42
Download 42 pages 61.46 Kb
Page 17
Image 17
Cal 0

5.At step 3, pressing the PRINT key to weigh the zero-point value. Do not touch the pan during weighing.

With nothing on the pan

To store

6.Place the calibration weight with the same value as displayed on the pan. Press the PRINT key to weigh it. Do not touch the pan during weighing.

7.

end

appears.

Remove the weight from the pan.

When the GLP output (function setting “info 1” or

info 2”) is selected,

glp

 

is displayed. The

calibration report is output and

 

end

appears

again.

 

 

 

 

Press the CAL switch or

UNITS

 

key to return to

the weighing mode.

 

 

 

 

Cal 0

50000.

Calibration weight

To store

50000.

0end000

glp

GLP output

0end000

Returns to

the weighing mode

‰Note

The value set in step 4 is stored in memory even after the power is switched off.

If the balance is to be moved to other places, set the gravity acceleration value of the area where the calibration using a weight is to be done, and calibrate the balance according to the procedure above. See the next section to set the value.