A&D EJ-6100, EJ-440, EJ-3000, EJ-200, EJ-120 2000, 1751, 8801, Place a sample on the upper pan

Models: EJ-4100

1 42
Download 42 pages 61.46 Kb
Page 31
Image 31
2.Place a sample on the upper pan.

2.Place a sample on the upper pan.

3.Wait for the STABLE indicator to be displayed and press the SAMPLE key to store the weight in air.

4.The display shows that LO blinks and the balance starts to measure weight in water.

5.Place the sample on the lower pan in water.

‰Adjust the amount of water so that the sample is about 10 mm below water surface.

6.Wait for the STABLE indicator to be displayed and press the SAMPLE key. Then the balance reads the weight in water and shows the density (specific gravity) of the sample.

7.To continue the specific gravity measurement, press the SAMPLE key again. To exit this measurement, press the UNITS key.

HI SG

2000.

Measuring 2000. weight in water.

LO SG

2000.

LO SG

1751.

SG

8801.1751.

The display shows the density.

To weigh in air.

To weigh in water.8801.

29

Page 31
Image 31
A&D EJ-6100, EJ-440, EJ-3000, EJ-4100, EJ-120, EJ-2000, EJ-1500 manual 1751, 8801, Place a sample on the upper pan