Service

5. Desolder the three leads of the backup battery (A2B1) with a solder removal tool and remove the battery.

6a. If the instrument has serial prefix 2942A and below, put the three insulators (HP 3050-0990)on the leads of the new battery (A2B1, HP 1420-0281)and solder in the battery. The bottom of the battery case must not touch the circuit board or any solder pads.

6b. If the instrument has serial prefix 2948A and above, solder in the battery (A2B1, HP 1420-0338).

7.Remove the resistor and temporary backup connections on the A2 assembly.

8.Re-install the A2 assembly and all connectors.

9.Place the instrument in its normal operating position. Re-insert the power cord and switch LINE to ON.

10.Press SHIFT PRESET. The display should show the correct instrument serial number suffix for a few seconds.

11.Re-install the instrument’s top cover.

rev.15DEC89

Troubleshooting (Memory Backup Battery)

8-48.3