Model 8904A

Service

3b. Clock Checks (Using a Counter)

a.If the counter has an external 10 MHz timebase reference input, connect it to the rear-panel 10 MHz REF OUT of the Multifunction Synthesizer.

NOTE

If the counter does not have an external reference, the readings in the following steps may be off several counts. This does not indicate a failure.

b.Connect the high-impedance counter input to pin 3 of U2. The counter should read within one count of 6.71088 MHz.

c.Connect the counter to pin 2 of U2. The counter should read within one count of 1.67772 MHz.

4a. Digital-to-AnalogConverter Check (Using an Oscilloscope)

a. On the Multifunction Synthesizer, press SHIFT PRESET. After the instrument presets, key in

For A3 Output Assembly For A10 Output Assembly

fl (Channel Config.)

fl (Channel Config.)

NEXT

NEXT NEXT

WAVE FORM (t

WAVE FORM (t

FREQ 25.6 Hz

FREQ 25.6 Hz

AMPTD 10 V

AMPTD 10 V

NOTE

Setting the frequency to exactly 25.6 Hz inputs a TTL squarewave to each input of the DAC latches. As the DAC’s data input lines are probed from the most significant bit the the least significant bit, the frequency is consecutively divided by two.

b.Connect a dc coupled, high-impedance oscilloscope to the pin of the IC listed below. The waveform should be a TTL squarewave with an approximate period as indicated in the table.

IC on A3 or A10

Pin

Period

u5

14

19.1 ps

u5

13

38.1 ps

u5

11

76.3 ps

u5

6

153 ps

u5

4

305 ps

u5

3

610 ps

u4

14

1.22 ms

u4

13

2.44 ms

u4

11

4.88 ms

u4

6

9.8 ms

u4

4

19.5 ms

u4

3

39.1 ms

Troubleshooting (Digital/Output Interface) 8-33