Page 117
Image 117

Service

Model 8904A

NOTE

a.If a typing error occurs as the Codeword is being entered, the backspace (em) key can be used to back up to the digit in error.

b.If the Codeword is incorrect the display will not list the desired option and after the 10 second timeout, the display will read:

* HP 8904A Main Selection Level * fl Channel Config.

Begin again at step 3.

6.Switch LINE to OFF and then back to ON. After the instrument powers up, the display should again read

* HP 89046 Main Selection Level * fl Channel Config.

7.Press SHIFT PRESET. Confirm that the option has been restored. The displaywill look something like following. (The example shows that Option 003 has been installed.)

I HP 8904A Opts 02/01/03

1

Firmware Revision 18387A Serial No 12345

I

8. If another option is to be restored in the same instrument, start again at step 3.

8-48Troubleshooting (Options 001 and 003)

Page 117