28 Language Dictionary
*RCL
Meaning and Type
Recall Device State
Recalling a previously stored state may place hazardous voltage at the power module output.
Description
This command restores the power module to a state that was previously stored in memory with a *SAV command to the
specified location. The following states are recalled:
CAL:AUTO LIST:COUN OUTP:REL[:STAT] TRIG:LINK
CURR[:LEV][:IMM] LIST:STEP OUTP:REL:POL TRIG:SOUR
CURR:MODE OUTP[:STAT] OUTP:TTLT[:STAT] VOLT[:LEV][IMM]
CURR:PROT:STAT OUTP:DFI[:STAT] OUTP:TTLT:LINK VOLT:MODE
DISP:STAT OUTP:DFI:LINK OUTP:TTLT:SOUR VOLT:PROT[:LEV]
INIT:CONT OUTP:PROT:DEL TRIG:DEL
Sending *RCL also does the following:
Forces an ABORt command before resetting any parameters (this cancels any uncompleted trigger actions).
Disables the calibration function by setting CAL:STATe to OFF.
The device state stored in location 0 is automatically recalled at power turn-on when the power module configuration switch
is set for this mode of operation (see the power module User’s Guide).
Note Whenever the power module is powered up, the state stored in location 0 is written to the 5 volatile
locations (5 through 9).
Command Syntax *RCL <NRf>
Parameters 0 through 9
Example *RCL 3
Query Syntax (None)
Related Commands *PSC *RST *SAV
*RST
Meaning and Type
Reset Device State
Description
This command resets the power module to a factory-defined state as defined below. *RST also forces an ABORt command.
COMMAND STATE COMMAND STATE COMMAND STATE
CAL:AUTO OFF OUTP[:STAT] OFF OUTP.TTLT:LINK OFF
CAL:STAT OFF OUTP:DFI OFF TRIG:DEL 0
CURR[:LEV][:IMM] 1OUTP:DFI:SOUR LINK TRIG:LINK OFF
CURR:PROT:STAT OFF OUTP:DFI:LINK “SUM3” TRIG:SOUR BUS
CURR:MODE FIX OUTP:PROT:DEL 1VOLT[:LEV][:IMM] 0
DISP[:WIND]:STAT ON OUTP:REL[:STAT] OFF VOLT:MODE FIX
INIT:CONT OFF OUTP:REL:POL NORM VOLT:PROT:LEV MAX
LIST:STEP A UTO OUTP:TTLT[STAT] OFF 1Model-dependent value. See
LIST:COUN 1OUTP:TTLT:SOUR BUS Table 3-2.