Language Dictionary 47
Command Syntax TRIGger[:STARt][:IMMediate]
Parameters (None)
Examples TRIG TRIG: IMM
Query Syntax (None)
Related Commands ABOR CURR:MODE CURR:TRIG INIT *TRG
VOLT:MODE VOLT:TRIG
TRIG:DEL
This command sets the time delay between the detection of an event on the specified trigger source and the start of any
corresponding trigger action on the power module’s output.
Command Syntax TRIGger[:STARt]:DELay <NRf+>
Parameters 0 to 65 | MIN | MAX
Default Suffix S
*RST Value 0
Examples TRIG:DEL .25 TRIG:DEL MAX
Query Syntax TRIGger[:STARt]:DELay?
Returned Parameters <NR3>
Related Commands ABOR CURR:TRIG INIT *TRG TRIG[:IMM]
VOLT:TRIG
TRIG:LINK
This command specifies which event conditions within the power module are linked to trigger source events when LINK is
the parameter of the TRIG:SOUR command.
Command Syntax TRIGger[:STARt]:LINK <CRD>
Parameters (See Table 3-1)
*RST Value OFF
Examples TRIG:LINK "CC" TRIG:LINK "OPER"
Query Syntax TRIGger[:STARt]:LINK?
Returned Parameters <CRD> (See Table 3-1)
Related Commands ABOR INIT *TRG TRIG[:IMM] TRIG:SOUR
TRIG:SOUR
This command selects the power module input trigger source as follows:
BUS *TRG or <GET> (Group Execute Trigger)
EXT Mainframe backplane Trigger In bus
HOLD No trigger source except TRIG:IMM
LINK Internal power module event as specified by TRIG:LINK
TTLT Mainframe Trigger Out bus
Command Syntax TRIGger[:STARt]:SOURce <CRD>
Parameters BUS | EXT | HOLD | LINK | TTLT
*RST Value BUS
Examples TRIG: SOUR BUS TRIG: SOUR LINK
Query Syntax TRIGger[:STARt]:SOURce?
Returned Parameters BUS | EXT | HOLD | TTLT
Related Commands ABOR CURR:TRIG INIT OUTP:TTLT
VOLT:TRIG