Command Syntax

TRIGger[:STARt][:IMMediate]

Parameters

(None)

Examples

TRIG TRIG: IMM

Query Syntax

(None)

Related Commands

ABOR CURR:MODE CURR:TRIG INIT *TRG

 

VOLT:MODE VOLT:TRIG

TRIG:DEL

This command sets the time delay between the detection of an event on the specified trigger source and the start of any corresponding trigger action on the power module’s output.

Command Syntax

TRIGger[:STARt]:DELay <NRf+>

Parameters

0 to 65 MIN MAX

Default Suffix

S

*RST Value

0

Examples

TRIG:DEL .25 TRIG:DEL MAX

Query Syntax

TRIGger[:STARt]:DELay?

Returned Parameters

<NR3>

Related Commands

ABOR CURR:TRIG INIT *TRG TRIG[:IMM]

 

VOLT:TRIG

TRIG:LINK

This command specifies which event conditions within the power module are linked to trigger source events when LINK is the parameter of the TRIG:SOUR command.

Command Syntax

TRIGger[:STARt]:LINK <CRD>

Parameters

(See Table 3-1)

 

*RST Value

OFF

 

Examples

TRIG:LINK "CC"

TRIG:LINK "OPER"

Query Syntax

TRIGger[:STARt]:LINK?

Returned Parameters

<CRD> (See Table 3-1)

Related Commands

ABOR INIT

*TRG TRIG[:IMM] TRIG:SOUR

TRIG:SOUR

This command selects the power module input trigger source as follows:

BUS *TRG or <GET> (Group Execute Trigger)

EXT Mainframe backplane Trigger In bus

HOLD No trigger source except TRIG:IMM

LINK Internal power module event as specified by TRIG:LINK

TTLT Mainframe Trigger Out bus

Command Syntax

TRIGger[:STARt]:SOURce <CRD>

Parameters

BUS EXT HOLD LINK TTLT

*RST Value

BUS

Examples

TRIG: SOUR BUS TRIG: SOUR LINK

Query Syntax

TRIGger[:STARt]:SOURce?

Returned Parameters

BUS EXT HOLD TTLT

Related Commands

ABOR CURR:TRIG INIT OUTP:TTLT

 

VOLT:TRIG

Language Dictionary 47

Page 46
Image 46
Agilent Technologies 66lxxA manual Trig Trig IMM, Trigdel .25 Trigdel MAX, Triglink Oper, Trigsour