Command Syntax

OUTPut[:STATe] <bool>[,NORelay]

Parameters

0

OFF[,NORelay] 1 ON[,NORelay]

*RST Value

0

 

 

Examples

OUTP 1

OUTP:STAT ON,NORELAY

Query Syntax

OUTPut[:STATe]?

Returned Parameters

0

1

 

Related Commands

*RCL

*SAV

OUTP:DFI

This command enables or disables the discrete fault indicator (DFI) signal to the power module backplane.

Command Syntax

OUTPut:DFI[:STATe] <bool>

Parameters

0 1 OFF ON

*RST Value

OFF

Examples

OUTP:DFI:1 OUTP:DFI OFF

Query Syntax

OUTPut:DFI[:STATe]?

Returned Parameters

0 1

Related Commands

OUTP:DFI:LINK OUTP:DFI:SOUR

OUTP:DFI:LINK

This command specifies which events within the power module are linked to DFI source events.

Command Syntax

OUTPut:DFI:LINK <CRD>

Parameters

See Table 3-2

 

*RST Value

SUM3

 

Examples

OUTP:DFI:LINK "CC"

OUTP:DFI:LINK “OFF”

Query Syntax

OUTPut:DFI:LINK?

Returned Parameters

See Table 3-2

 

Related Commands

OUTP:DFI:SOUR

OUTP:DFI[:STAT]

OUTP:DFI:SOUR

This command selects the source for DFI events. The only available source is LINK.

Command Syntax

OUTP:DFI:SOUR <CRD>

Parameters

LINK

*RST Value

LINK

Examples

OUTP:DFI:SOUR LINK

Query Syntax

OUTPut:DFI:SOUR?

Returned Parameters

LINK

Related Commands

OUTP:DFI:LINK OUTP:DFI[:STAT]

OUTP:PROT

There are two output protection commands that do the following:

OUTP:PROT:CLE Clears any overvoltage (OV), overcurrent (OC), overtemperature (OT), or remote inhibit (RI) protection features. After this command, the output is restored to the state it was in before the protection feature occurred.

40 Language Dictionary

Page 39
Image 39
Agilent Technologies 66lxxA manual Outp Outpstat ON,NORELAY, Outpprot