40 Language Dictionary
Command Syntax OUTPut[:STATe] <bool>[,NORelay]
Parameters 0 | OFF[,NORelay] | 1 | ON[,NORelay]
*RST Value 0
Examples OUTP 1 OUTP:STAT ON,NORELAY
Query Syntax OUTPut[:STATe]?
Returned Parameters 0 | 1
Related Commands *RCL *SAV
OUTP:DFI
This command enables or disables the discrete fault indicator (DFI) signal to the power module backplane.
Command Syntax OUTPut:DFI[:STATe] <bool>
Parameters 0 |1 | OFF | ON
*RST Value OFF
Examples OUTP:DFI:1 OUTP:DFI OFF
Query Syntax OUTPut:DFI[:STATe]?
Returned Parameters 0 | 1
Related Commands OUTP:DFI:LINK OUTP:DFI:SOUR
OUTP:DFI:LINK
This command specifies which events within the power module are linked to DFI source events.
Command Syntax OUTPut:DFI:LINK <CRD>
Parameters See Table 3-2
*RST Value SUM3
Examples OUTP:DFI:LINK "CC" OUTP:DFI:LINK “OFF”
Query Syntax OUTPut:DFI:LINK?
Returned Parameters See Table 3-2
Related Commands OUTP:DFI:SOUR OUTP:DFI[:STAT]
OUTP:DFI:SOUR
This command selects the source for DFI events. The only available source is LINK.
Command Syntax OUTP:DFI:SOUR <CRD>
Parameters LINK
*RST Value LINK
Examples OUTP:DFI:SOUR LINK
Query Syntax OUTPut:DFI:SOUR?
Returned Parameters LINK
Related Commands OUTP:DFI:LINK OUTP:DFI[:STAT]
OUTP:PROT
There are two output protection commands that do the following:
OUTP:PROT:CLE Clears any overvoltage (OV), overcurrent (OC), overtemperature (OT), or remote inhibit (RI)
protection features. After this command, the output is restored to the state it was in before the
protection feature occurred.