46 Language Dictionary
Command Syntax STATus:QUEStionable:NTRansition <NRf>
STATus:QUEStionable:PTRansition <NRf>
Parameters 0 to 32727
Suffix (None)
Default Value 0
Example STAT:QUES:NTR 16 STAT:QUES:PTR 512
Query Syntax STATus:QUEStionable:NTRansition?
STATus:QUEStionable:PTRansitiion?
Returned Parameters <NR1> (Register value)
Related Commands STAT:QUES:ENAB
SYST:ERR?
This query returns the next error number followed by its corresponding error message string from the remote programming
error queue. The queue is a FIFO (first-in, first-out) buffer that stores errors as they occur. As it is read, each error is
removed from the queue. When all errors have been read, the query returns 0,NO ERROR. If more errors are accumulated
than the queue can hold, the last error in the queue is -350,TOO MANY ERRORS.
Query Syntax SYSTem:ERRor?
Parameters (None)
Returned Parameters <NR1>,<SRD>
Example SYST:ERR?
SYST:VERS?
This query returns the SCPI version number to which the power module complies. The returned value is of the form
YYYY.V, where YYYY represents the year and V is the revision number for that year.
Query Syntax SYSTem:VERSion?
Parameters (none)
Returned Parameters <NR2>
Example SYST:VERS?
Related Commands (None)
Trigger Subsystem
This subsystem controls the triggering of the power module. See "Chapter 5 - Synchronizing Power Module Output
Changes" for an explanation of the Trigger Subsystem.
Note The trigger subsystem must be enabled from the Initiate Subsystem or no triggering action will occur.
TRIG
When the trigger subsystem is enabled, TRIG generates an immediate trigger signal that bypasses any selected
TRIG:SOUR and TRIG:DEL. The trigger will then:
1. Initiate a pending level change as specified by CURR[:LEV]:TRIG or VOLT[:LEV]:TRIG.
2. Initiate a pending level change as specified by CURR:MODE LIST or VOLT:MODE LIST and in accordance
with LIST:STEP.
3. Clear the WTG bit in the Status Operation Condition register.