OUTP:PROT:DEL Sets the delay time between the programming of an output change that produces a CV, CC, or UNREG condition and the recording of that condition by the Status Operation Condition register. The delay prevents momentary changes in power module status that can occur during reprogramming from being registered as events by the status subsystem. Since the delay applies to CC status, it also delays the OCP (overcurrent protection) feature. The OVP (overvoltage protection) feature is not affected by this delay.

Command Syntax

OUTPut:PROTection:CLEar

 

OUTPut:PROTection:DELay <NRf+>

Parameters

OUTP:PROT:CLE,

(none)

 

OUTP:PROT:DEL

0 to 32.767 MIN MAX

Default Suffix

S

 

*RST Value

100 (milliseconds)

 

Examples

OUTP:PROT:CLE OUTP:PROT:DEL 75E-1

Query Syntax

OUTP:PROTection:CLEar (None)

 

OUTPut:PROTection:DELay?

 

OUTPut:PROTection:DELay? MINimum

 

OUTPut:PROTection:DELay? MAXimum

Returned Parameters

<NR3>

 

Related Commands

OUTP:PROT:CLE (None)

 

OUTP:PROT:DEL

*RCL *SAV

OUTP:REL

This command is valid only if the power module is configured for the optional relay connector. Programming ON closes the relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the power module is supplying power to a load, that power will appear at the relay contacts during switching. If the power module is not configured for the relay connector, sending either relay command generates an error.

Command Syntax

OUTPut:RELay[:STATe] <bool>

Parameters

0

1 OFF ON

 

*RST Value

0

 

 

Examples

OUTP:REL 1

OUTP:REL OFF

Query Syntax

OUTPut:RELay?

Returned Parameters

0

1

 

Related Commands

OUTP[:STAT]

*RCL *SAV

OUTP:REL:POL

This command is valid only if the power module is configured for the optional relay connector. Programming NORMal causes the relay output polarity to be the same as the power module output. Programming REVerse causes the relay output polarity to be opposite to that of the power module output. If OUTP[:STAT] = ON when either relay command is sent, the power module output voltage is set to 0 during the time that the relays are changing polarity. If the power module is not configured for the relay connector, sending either relay command generates an error.

Command Syntax

OUTPut:RELay:POLarity <CRD>

Parameters

NORMal REVerse

*RST Value

NORM

 

Examples

OUTP:REL:POL

NORM

Query Syntax

OUTPut:RELay:POLarity?

Returned Parameters

NORM REV

 

Related Commands

OUTP[:STAT]

*RCL *SAV