Language Dictionary 41
OUTP:PROT:DEL Sets the delay time between the programming of an output change that produces a CV, CC, or
UNREG condition and the recording of that condition by the Status Operation Condition register.
The delay prevents momentary changes in power module status that can occur during
reprogramming from being registered as events by the status subsystem. Since the delay applies
to CC status, it also delays the OCP (overcurrent protection) feature. The OVP (overvoltage
protection) feature is not affected by this delay.
Command Syntax OUTPut:PROTection:CLEar
OUTPut:PROTection:DELay <NRf+>
Parameters OUTP:PROT:CLE, (none)
OUTP:PROT:DEL 0 to 32.767 | MIN | MAX
Default Suffix S
*RST Value 100 (milliseconds)
Examples OUTP:PROT:CLE OUTP:PROT:DEL 75E-1
Query Syntax OUTP:PROTection:CLEar (None)
OUTPut:PROTection:DELay?
OUTPut:PROTection:DELay? MINimum
OUTPut:PROTection:DELay? MAXimum
Returned Parameters <NR3>
Related Commands OUTP:PROT:CLE (None)
OUTP:PROT:DEL *RCL *SAV
OUTP:REL
This command is valid only if the power module is configured for the optional relay connector. Programming ON closes
the relay contacts; programming OFF opens them. The relay is controlled independently of the output state. If the power
module is supplying power to a load, that power will appear at the relay contacts during switching. If the power module is
not configured for the relay connector, sending either relay command generates an error.
Command Syntax OUTPut:RELay[:STATe] <bool>
Parameters 0 | 1 | OFF | ON
*RST Value 0
Examples OUTP:REL 1 OUTP:REL OFF
Query Syntax OUTPut:RELay?
Returned Parameters 0 |1
Related Commands OUTP[:STAT] *RCL *SAV
OUTP:REL:POL
This command is valid only if the power module is configured for the optional relay connector. Programming NORMal
causes the relay output polarity to be the same as the power module output. Programming REVerse causes the relay output
polarity to be opposite to that of the power module output. If OUTP[:STAT] = ON when either relay command is sent, the
power module output voltage is set to 0 during the time that the relays are changing polarity. If the power module is not
configured for the relay connector, sending either relay command generates an error.
Command Syntax OUTPut:RELay:POLarity <CRD>
Parameters NORMal | REVerse
*RST Value NORM
Examples OUTP:REL:POL NORM
Query Syntax OUTPut:RELay:POLarity?
Returned Parameters NORM | REV
Related Commands OUTP[:STAT] *RCL *SAV