50 Language DictionaryTable 3-1. Link Parameter List
Parameter True Event Condition Valid for
CC Constant current event bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
CV Constant voltage event bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
ESB Standard event summary bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
LSC List sequence complete pulse2OUTP:TTLT:LINK TRIG[:STAR]:LINK
OPC Operation complete bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
OPER Operation summary bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
QUES Questionable summary bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
MAV Message available summary bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
OFF No linked event condition OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
RQS Request service summary bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
RTG Received a trigger bit2OUTP:TTLT:LINK TRIG[:STAR]:LINK
STC List step completed pulse2OUTP:TTLT:LINK TRIG[:STAR]:LINK
STS List step started pulse2OUTP:TTLT:LINK TRIG[:STAR]:LINK
SUM3 OPER or QUES or ESB bit1OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK
TDC Trigger delay complete pulse 2OUTP:TTLT:LINK TRIG[:STAR]:LINK
1 See “Chapter 4 - Status Reporting” 2 See “Chapter 5 - Synchronizing Power Module Output Changes”.
Table 3-2. Power Module Programming Parameters
Agilent Model
Parameter 66101A 66102A 66103A 66104A 66105A 66106A
Output Programming Range
(maximum programmable values):
Voltage: 8.190 V 20.475 V 35.831 V 61.425 V 122.85 V 204.75 V
Current: 16.380 A 7.678 A 4.607 A 2.559 A 1.280 A 0.768 A
OV Protection: 8.8 V 22.0 V 38.5 V 66.0 V 132.0 V 220.0 V
Average Resolution
Voltage: 2.4 mV 5.9 mV 10.4 mV 18.0 mV 36.0 mV 60.0 mV
Current: 4.6 mA 2.3 mA 1.4 mA 0.75 mA 0.38 mA 0.23 mA
OV Protection: 50 mV 120 mV 200 mV 375 mV 750 mV 1.25 V
*RST State Values1
Voltage: 0
Current: 257 mA 120 mA 72 mA 40 mA 20 mA 12 mA
OV Protection: 8.8 V 22.0 V 38.5 V 66.0 V 132.0 V 220.0 V
1 These also are the power-on reset values when the factory default parameters are left in effect (see "Chapter 2
Installation” in the power module User’s Guide).Nonvolatile Memory Locations: 5; (0 through 4)Volatile Memory Locations: 5; (5 through 9):