Table 3-1. Link Parameter List

Parameter

CC

CV

ESB

LSC

OPC

OPER

QUES

MAV

OFF

RQS

RTG

STC

STS

SUM3

TDC

True Event Condition

Constant current event bit1 Constant voltage event bit1 Standard event summary bit1 List sequence complete pulse2 Operation complete bit1 Operation summary bit1 Questionable summary bit1 Message available summary bit1 No linked event condition Request service summary bit1 Received a trigger bit2

List step completed pulse2 List step started pulse2 OPER or QUES or ESB bit1 Trigger delay complete pulse 2

Valid for

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:DFI:LINK OUTP:TTLT:LINK TRIG[:STAR]:LINK

OUTP:TTLT:LINK TRIG[:STAR]:LINK

1 See “Chapter 4 - Status Reporting”

2 See “Chapter 5 - Synchronizing Power Module Output Changes”.

 

 

Table 3-2. Power Module Programming Parameters

Parameter

Output Programming Range (maximum programmable values):

66101A

66102A

Agilent Model

66103A

66104A

66105A

66106A

Voltage:

Current:

OV Protection:

Average Resolution

8.190 V

16.380 A

8.8 V

20.475 V

7.678 A

22.0 V

35.831 V

4.607 A

38.5 V

61.425 V

2.559 A

66.0 V

122.85V

1.280 A

132.0V

204.75V

0.768 A

220.0V

Voltage:

Current:

OV Protection:

*RST State Values1

Voltage:

Current:

OV Protection:

2.4mV

4.6mA 50 mV

257mA

8.8 V

5.9mV

2.3mA 120 mV

120mA

22.0 V

10.4 mV

 

18.0 mV

1.4 mA

 

0.75 mA

200 mV

 

375 mV

 

 

 

 

0

72 mA

 

40 mA

38.5 V

 

66.0 V

36.0mV

0.38mA 750 mV

20mA

132.0 V

60.0mV

0.23mA

1.25V

12mA

220.0 V

1These also are the power-on reset values when the factory default parameters are left in effect (see "Chapter 2 Installation” in the power module User’s Guide ).

Nonvolatile Memory Locations:

5; (0 through 4)

Volatile Memory Locations:

5; (5 through 9)

:

 