Agilent Technologies E1446A user manual 22 E1446/E1405/06 Command Reference SYSTem Subsystem

Models: E1446A

1 161
Download 161 pages 46.15 Kb
Page 101
Image 101

SYSTem

The SYSTem subsystem returns error messages and the SCPI version number to which the Agilent E1446A complies.

Subsystem Syntax SYSTem

 

:ERRor?

[query only]

:VERSion?

[query only]

:ERRor?

SYSTem:ERROR? returns the error messages in the error queue. See Appendix B for a listing of possible error numbers and messages.

Comments The Agilent E1446A places any generated errors into the error queue. The queue

is first-in, first out. With several errors waiting in the queue, the SYSTem:ERRor? returns the oldest unread error message first.

The error queue can hold 30 error messages. If the Agilent E1446A generates more than 30 messages that are not read, it replaces the last error message in the queue with error -350,"Too many errors". No additional messages are placed into the queue until SYSTem:ERRor? reads some messages or the *CLS (clear status) command clears the queue.

When the error queue is empty, SYSTem:ERRor? returns +0,"No error".

Executable when initiated: Yes

*RST Condition: unaffected

Power-On Condition: no errors are in the error queue

Example Reading the error queue

SYST:ERR?

Query the error queue

:VERSion?

SYSTem:VERSion? returns the SCPI version number to which the Agilent

E1446A complies: “1991.0”.

Comment Executable when initiated: Yes

*RST Condition: none

3-22 E1446/E1405/06 Command Reference

SYSTem Subsystem

Page 101
Image 101
Agilent Technologies E1446A user manual 22 E1446/E1405/06 Command Reference SYSTem Subsystem