OUTPut2 [:STATe]
Comments • A one (1) returned in response to the query indicates an overload condition exists. A zero (0) indicates there is not an overload condition.
•Coupling group: none
•*RST Condition: none
Example Determining if an overload condition exists
OUTP2:OVER? | determine if overload condition exists |
[:STATe]
OUTPut2[:STATe] <mode> closes or opens the output relay of the “Main Output” BNC to enable or disable the analog output. When disabled, the output appears as an open circuit.
Parameters |
| Parameter | Parameter | Range of | Default |
|
| ||||
|
| Name | Type | Values | Units |
|
|
|
|
|
|
|
| mode | boolean | OFF0ON1 | none |
Comments |
|
|
|
|
|
• Executable when initiated: Yes |
|
|
•Coupling group: none
•*RST Condition: OUTPut2:STATe ON
•The output relay will open automatically if an output overload is detected.
Example Disabling the output
OUTP2 OFF | Disable output |
[:STATe]:ACTual?
OUTPut2[:STATe]:ACTual? determines if the amplifier’s ’Main Output’ BNC is enabled or has been disabled due to an overload condition. The determination is done by reading bit 8 of the amplifier’s Status register (Appendix C).
This command requires Agilent E1445A firmware revision A.02.00 or greater.
OUTPut2 Subsystem | E1446/E1445 Command Reference |