*ESR?

Power-On Condition: no events are enabled

Example Enable all error events

*ESE 60

Enable error events

*ESR?

*ESR? returns the value of the Standard Event Status Register. The register is then cleared (all bits 0).

Comments Executable when initiated: Yes

Coupling group: none

*RST Condition: none

Power-On Condition: register is cleared

*GMC?

*GMC? <name> returns the definition of the specified macro in IEEE-488.2 definite block format.

Parameters

Parameter

Parameter

Range of

Default

 

 

Name

Type

Values

Units

 

 

 

 

 

 

name

string

defined macro name

none

 

 

data

 

 

 

 

 

 

 

Comments Executable when initiated: Yes

Coupling group: none

Related Commands: *DMC

*RST Condition: none

Power-On Condition: no macros are defined