Executing Coupled
Commands The list below identifies rules to follo w wh en exe c uting coupled commands:
Coupled commands must be contiguous and executed in t he same
program statement. This done by placing the commands in the same
program line, or by suppressing the end-of-line termina tor until the
last coupled comman d ha s been sent.
To send multiple commands in a single line or in a single statement,
the commands are linked (as described previously) with a semicolon
(;) and a colon(:). For example:
OUTP2:IMP 50;OUTP2:ATT 6
or
OUTP2:IMP 50;
:OUTP2:ATT 6
In BASIC programs, the end-of-line (EOL) terminator is suppr e sse d
by placing a semicolon (;) following the quotation ma rk (" ) wh ich
closes the command string. For example:
OUTPUT 70910;"OUTP2:IMP 50;";
OUTPUT 70910;":OUTP2:ATT 6"
OUTPUT 70910;"OUTP2:STAT ON"
As shown, the first two lines are coupled together. The third line is
not a coupled command, therefore, the EOL terminator is not
suppressed on the second line.
Commands not in the coupling group must either precee d or fo ll ow
commands in the coupling group.
Un-coupled commands executed in a coupling group break the
coupling.
Error checking occurs at the end of the coupling group.
Hardware updates occur at the end of the coupling group.