6 – Language Dictionary

CALibrate:DATA

This command enters a calibration value that you obtain by reading an external meter. You must first select a calibration level (with CALibrate:LEVel) for the value being entered.

Command Syntax

CALibrate:DATA <NRf>

 

Parameters

<external reading>

 

Unit

A or V (amperes or volts)

Examples

CAL:DATA

3222.3 MA

CAL:DATA 5.000

Related Commands

CAL:STAT

CAL:LEV

 

CALibrate:DATE

This command stores the date the unit was last calibrated. Enter any ASCII string up to 31 characters.

Command Syntax

CALibrate:DATE <date>

 

Parameters

<date>

 

Examples

CAL:DATE "3/22/01"

CAL:DATE "22.3.99"

Query Syntax

CALibrate:DATE?

 

Returned Parameters

<SRD>

 

CALibrate:LEVel

This command selects the next point in the calibration sequence. P1 is the first calibration point, P2 is the second calibration point.

Command Syntax

CALibrate:LEVel <point>

Parameters

P1 P2

Examples

CAL:LEV P2

CALibrate:PASSword

This command lets you change the calibration password. A new password is automatically stored in nonvolatile memory and does not have to be stored with CALibrate:SAVE. The default password is the model number of the unit. If the password is set to 0, password protection is removed and the ability to enter the calibration mode is unrestricted.

Command Syntax CALibrate:PASSword <NRf>

Parameters <model number> (default)

Examples CAL:PASS 1234

Related Commands CAL:SAV