B - Performance and Calibration Procedures

Performance Test Record Form

Model Agilent N3280A - Output 1

Report No ______________

Date __________________

Test Description

 

Minimum

Results

Maximum

 

 

 

Specification

 

 

Specification

 

 

VOLTAGE PRIORITY TESTS

 

 

Programming Accuracy (DMM readings)

 

 

 

 

 

 

 

Voltage ( 0V)

 

− 2mV

_________

+ 2mV

Voltage (+10V)

 

9.988 V

_________

10.012 V

Voltage (-10V)

 

9.988 V

_________

10.012 V

+ 1mA Current limit

 

0.949mA

_________

1.051mA

+ 0.5A Current limit

 

0.49945 A

_________

0.50055 A

− 1mA Current limit

 

−0.949mA

_________

− 1.051mA

− 0.5A Current limit

 

0.49945 A

_________

0.50055 A

Readback Accuracy (MEAS? readings)

 

 

 

 

 

 

 

Voltage ( 0V)

 

− 2mV

_________

+ 2mV

Voltage (+10V)

 

Vout − 12mV

_________

Vout + 12mV

Voltage (-10V)

 

Vout − 12mV

_________

Vout + 12mV

+ 0.5A range current

 

Iout − 0.7mA

_________

Iout + 0.7mA

− 0.5A range current

 

Iout − 0.7mA

_________

Iout + 0.7mA

+ 15mA range current

 

Iout

_________

 

15mA range current

 

15 A

_________

Iout + 15 A

 

Iout

 

 

 

 

15 A

_________

Iout + 15 A

+ 0.5mA range current

 

Iout − 0.7A

Iout + 0.7A

0.5mA range current

 

Iout

_________

 

 

 

 

0.7 A

 

 

Iout + 0.7 A

Load Effect

 

 

 

 

 

 

 

 

Voltage

 

− 400V

_________

+ 400V

+ Current limit

 

 

_________

+ 30∝A

− Current limit

 

30 A

_________

+ 30∝A

 

 

 

 

 

 

30 A

 

 

 

 

Source Effect

 

 

 

 

 

 

 

 

Voltage

 

− 200V

_________

− 200V

+ Current limit

 

 

_________

− Current limit

 

10 A

_________

10 A

 

 

 

 

 

 

10 A

 

 

 

10 A

PARD (Ripple and Noise)

 

 

 

 

 

 

 

 

Voltage (rms)

 

 

 

 

_________

380V

Voltage (peak-to-peak)

 

 

 

 

_________

4mV

±Current limit (rms)

 

 

 

 

_________

40∝A

Transient Response Time

Low( 10kHz):

 

 

 

_________

60∝s

 

 

Med ( 20kHz):

 

 

 

_________

45∝s

 

 

High ( 30kHz):

 

 

 

_________

35∝s

 

 

CURRENT PRIORITY TESTS

 

 

Programming Accuracy (DMM readings)

1∝A

 

 

 

 

Current ( 0A )

 

_________

+ 1∝A

Current ( 0.5mA)

 

0.0004985 A

_________

0.0005015 A

Current ( −0.5mA)

 

0.0004985 A

_________

0.0005015 A

Load Effect

 

 

 

 

 

 

 

 

Current

 

− 25nA

_________

+ 25nA

Source Effect

 

 

 

 

 

 

 

 

Current

 

− 10nA

_________

+ 10nA

PARD (Ripple and Noise)

 

 

 

 

 

 

 

 

Current (rms)

 

 

 

 

_________

1.5∝A

Transient Response Time

 

 

 

 

_________

 

90∝s