Performance and Calibration Procedures

3.

Read back the N3280A status to be sure that it’s in the CLmode. This

“STAT:OPER:COND? (@1)”

 

query should return a Bit value of “4” for CLmode.

 

 

4.

Adjust the transformer to the lowest rated line voltage

 

 

(e.g., 104 Vac for a 120 Vac nominal line voltage input).

 

 

Set the ammeter to the 1A range, and record the current reading on the

 

 

ammeter. (low-line value)

 

5.

Adjust the transformer to the highest rated line voltage

 

 

(e.g., 127 Vac for 120 Vac nominal line voltage input).

 

 

Record the current reading on the ammeter. (high-line value)

 

6.

The difference between the low-line and the high-line values is the

 

 

source effect voltage and should be within the limits listed in the

 

 

performance test record card under Voltage Priority Source Effect

 

 

–Current Limit.

 

Current Priority, Constant Current Source Effect

NOTE: The voltage limits in Current Priority Mode are not programmable.

Action

Program Commands

1.Connect the ac input of the dc source to a variable voltage transformer (or ac source). Set the transformer to nominal line voltage.

Connect the output as shown in Figure B-1a with an ammeter directly across the Hi and Lo output terminals.

2.Turn on the dc source and program the Current Priority mode. Program the current to 0.5mA.

3.Read back the N3280A status to be sure that it’s in the CC mode. This query should return a Bit value of “8” for CC mode.

4.Adjust the transformer to the lowest rated line voltage (e.g., 104 Vac for a 120 Vac nominal line voltage input).

Set the ammeter to the 1mA range, and record the current reading on the ammeter. (low-line value)

5.Adjust the transformer to the highest rated line voltage (e.g., 127 Vac for 120 Vac nominal line voltage input). Record the current reading on the ammeter. (high-line value)

6.The difference between the low-line and the high-line values is the source effect current and should be within the limits listed in the performance test record card under Current Priority Source Effect Current.

“OUTP ON,(@1)” “SOUR:FUNC:MODE CURR,(@1)” “CURR 0.0005,(@1)”

“STAT:OPER:COND? (@1)”

93

Page 93
Image 93
Agilent Technologies N3280A manual Current Priority, Constant Current Source Effect