Language Dictionary - 6

[SOURce:]CURRent[:IMMediate]

[SOURce:]CURRent:TRIGgered

These commands set the immediate and the pending triggered current level of the dc source. They only apply in current priority mode. The immediate level is the output current setting. The pending triggered level is a stored value that is transferred to the output when a trigger occurs. To respond to a trigger, the [SOUR:]CURR:MODE must be set to STEP, and the trigger system must be initiated.

Command Syntax

[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude] <Nrf+>, (@<channel list>)

 

[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude] <Nrf+>, (@<channel list>)

Parameters

see table 6-3

 

Default Suffix

A ( amperes)

 

*RST Value

0

 

Examples

CURR 0.0001 ,(@1)

CURR:TRIG 0.0002,(@1)

Query Syntax

[SOURce:]CURRent[:LEVel][:IMMediate][:AMPLitude]? (@<channel list>)

 

[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]? (@<channel list>)

Returned Parameters

<NR3>

 

Related Commands

INIT CURR:MODE

 

[SOURce:]CURRent:LIMit[:IMMediate]

[SOURce:]CURRent:LIMit:TRIGgered

These commands set the immediate and the pending triggered current limit of the dc source. They only apply in voltage priority mode. The current limit setting applies to both the positive and negative current limits. The pending triggered limit is a stored value that applies when a trigger occurs. To respond to a trigger, the [SOUR:]CURR:LIM:MODE must be set to STEP, and the trigger system must be initiated.

Command Syntax

[SOURce:]CURRent:LIMit[:POSitive][:IMMediate] <Nrf+>, (@<channel list>)

 

[SOURce:]CURRent:LIMit[POSitive]:TRIGgered <Nrf+>, (@<channel list>)

Parameters

see table 6-3

 

Default Suffix

A ( amperes)

 

*RST Value

0.001

 

Examples

CURR:LIM 0.25,(@1)

CURR:LIM:TRIG 0.35,(@1)

Query Syntax

[SOURce:]CURRent:LIMit[:POSitive][:IMMediate]? (@<channel list>)

 

[SOURce:]CURRent:LIMit[POSitive]:TRIGgered? (@<channel list>)

Returned Parameters

<NR3>

 

Related Commands

INIT CURR:LIM:MODE

 

[SOURce:]CURRent:LIMit:BWIDth

This command configures the output compensation band of the current limit circuit. If capacitive loads cause the output to oscillate, use this command to select a lower compenstion band. Note that if the output is on, changing the compensation will cause the output to cycle OFF, then ON. The following compensation bandwidths may be programmed: 30 kHz or 10 kHz.

Command Syntax

[SOURce:]CURRent:LIMit:BWIDth <Nrf>, (@<channel list>)

Parameters

10000 30000

*RST Value

30000

Examples

CURR:LIM:BWID 10000,(@1)

Query Syntax

[SOURce:]CURRent:LIMit:BWIDth? (@<channel list>)

Returned Parameters

<NR3>

65

Page 65
Image 65
Agilent Technologies N3280A manual SOURceCURRentIMMediate SOURceCURRentTRIGgered, SOURceCURRentLIMitBWIDth, Default Suffix