Agilent Technologies N3280A manual SOURceVOLTageMODE, SOURceVOLTagePROTectionSTATe, NR10 or

Models: N3280A

1 116
Download 116 pages 15.96 Kb
Page 68
Image 68

6 – Language Dictionary

[SOURce:]VOLTage:MODE

This command determines what happens to the output voltage during a triggered event.

FIXed

The output voltage is unaffected when a trigger occurs.

STEP

The output voltage is programmed to the value set by VOLT:TRIG when a trigger occurs.

Command Syntax

[SOURce:]VOLTage:MODE <mode>, (@<channel list>)

 

Parameters

FIXed STEP

 

*RST Value

FIXed

 

Examples

VOLT:MODE FIX,(@1)

 

Query Syntax

[SOURce:]VOLTage:MODE? (@<channel list>)

Returned Parameters

<CRD>

Related Commands

VOLT:TRIG

[SOURce:]VOLTage:PROTection:STATe

This command enables or disables the overvoltage protection (OVP) function. The command only applies in voltage priority mode. When enabled, the output of the unit will shut down and the output relays will open when the output voltage exceeds +11.5V (±0.3V), or −11.5V (±0.3V).

CAUTION: Disabling the overvoltage protection function may cause excessive output voltages, such as can occur if remote sense leads are shorted, to damage the equipment under test.

Command Syntax

[SOURce:]VOLTage:PROTection:STATe <bool>, (@<channel list>)

Parameters

0 OFF 1 ON

*RST Value

ON

Examples

VOLT:PROT:STAT 0,(@1)

Query Syntax

[SOURce:]VOLTage:PROTection:STATe? (@<channel list>)

Returned Parameters

<NR1>0 or 1

68

Page 68
Image 68
Agilent Technologies N3280A manual SOURceVOLTageMODE, SOURceVOLTagePROTectionSTATe, NR10 or