Language Dictionary - 6

SENSe:SWEep:POINts

This command defines the number of points in a measurement.

Command Syntax

SENSe:SWEep:POINts <NRf+>

Parameters

1 through 4096

*RST Value

5

Examples

SENS:SWE:POIN 1024

Query Syntax

SENSe:SWEep:POINts?

Returned Parameters

<NR3>

Related Commands

SENS:SWE:TINT SENS:SWE:OFFS MEAS:ARR

SENSe:SWEep:TINTerval

This command defines the time period between samples. The value that you enter for the time interval will be rounded to the nearest 30.4 microsecond increment.

Command Syntax

SENSe:SWEep:TINTerval <NRf+>

Parameters

30.4 microseconds through 60800 seconds

*RST Value

30.4 microseconds

Examples

SENS:SWE:TINT 60.8E-6

Query Syntax

SENSe:SWEep:TINTerval?

Returned Parameters

<NR3>

Related Commands

SENS:SWE:POIN SENS:SWE:OFFS MEAS:ARR

SENSe:WINDow

This command sets the window function that is used in dc measurement calculations. The following functions can be selected:

HANNing A signal conditioning window that reduces errors in dc measurement calculations in the presence of periodic signals such ac line ripple. The Hanning window multiplies each point in the measurement sample by the function cosine4.

RECTangular A window that returns measurement calculations without any signal conditioning.

NOTE: Neither window function alters the instantaneous voltage or current data returned in the measurement array.

Command Syntax SENSe:WINDow[:TYPE] <type> Parameters HANNing RECTangular *RST Value RECTangular

Examples SENS:WIND RECT

Query Syntax SENSe:WINDow[:TYPE]?

Returned Parameters <CRD>

63

Page 63
Image 63
Agilent Technologies N3280A manual SENSeSWEepPOINts, SENSeSWEepTINTerval, SENSeWINDow