B - Performance and Calibration Procedures

Table B-1. Equipment Required (continued)

Oscilloscope

Sensitivity: 1 mV/div.

Agilent Infinium or

 

Bandwidth Limit: 20 to 30 MHz

equivalent

 

Probe: 1:1 with RF tip

 

RMS voltmeter

True RMS

Rhode & Schwartz

 

Bandwidth: 20 Mhz min.

Model URE3 RMS-P-P

 

Sensitivity: 100 ∝V

Voltmeter

Variable-voltage transformer

Adjustable to highest rated input voltage

Agilent 6800 series

or ac source

range. Power: 500 VA

 

Tektronixs current probe

20mA/div

AM503B

amplifier and power module

 

TM501 or 2A

Pulse/function generator

+/- 1V Square Wave , 400-1kHz

Agilent 8116A

Load resistor, 20 ohms

20 ohm 10W

0811-3896 or equivalent

Resistor 0.2 ohm

5- 1 ohm resistors in parallel

5 – 0699-0208 or equivalent

Resistor 1k ohm

1k ohm for 15mA range accuracy

0757-0280 or equivalent

Resistor 20K ohm

20k ohm for 0.5mA range accuracy

0757-0449 or equivalent

Resistor 16k ohm

16.2k ohm for rms noise measurements

0757-0447 or equivalent

Resistor 50 ohm

50 ohm series resistor for noise

0757-0706 or equivalent

 

measurements

 

Capacitor 10uF for voltage

3- 3.3uF film type capacitors in parallel

3 – 0160-7308 or equivalent

transient response

 

 

Capacitor for current

0.47uF film type capacitor

1060-0970 or equivalent

transient response

 

 

Performance & Verification Tests

Enter all of the performance test results and calculated measurements in the Performance Test Record Form that is provided at the end of this section.

Measurement Techniques

If more than one meter or if a meter and an oscilloscope are used, connect each to the terminals by a separate pair of leads to avoid mutual coupling effects. For constant voltage dc tests, connect only to HI sense and LO sense terminals, since the unit regulates the voltage that appears at the sense terminals, not the output terminals. Use twisted-pair wiring to avoid noise pickup on the test leads.

NOTE: When using the Agilent 3458A as an ammeter, always select the specific current measurement range that you will be using. Do not use the autoranging feature of the ammeter, as this may introduce noise in your current measurements by toggling between measurement ranges. Always use the lowest range possible to provide the best measurement accuracy.