Performance Tests and Calibration

Introduction

This appendix contains test procedures to verify that the dc source is operating normally and is within published specifications. There are three types of tests as follows:

Built-in Self Tests

These tests run automatically when the dc source is turned on. They

 

check most of the digital circuits and the programming and readback

 

DACs.

Turn on Checkout

These tests, described in chapter four, provide a high degree of

 

confidence that your unit is operating properly.

Calibration Verification

These tests, documented in this appendix, verify that the dc source is

/Performance Tests

properly calibrated, and that the dc source meets all of the specifications

 

listed in Appendix A

If the dc source fails any of the tests or if abnormal test results are obtained after performing a calibration, return the unit to an Agilent Technologies repair facility.

This appendix also includes calibration procedures for the Agilent N3280A. Instructions are given for performing the procedures from a controller over the GPIB.

IMPORTANT: Perform the Programming Accuracy and Readback Accuracy tests before calibrating your dc source. If the dc source passes the Programming Accuracy and Readback Accuracy tests, the unit is operating within its calibration limits and does not need to be re-calibrated.

Equipment Required

The equipment listed in the following table, or the equivalent to this equipment, is required for the calibration and performance tests. A test record sheet with specification limits (when test using the recommended test equipment) may be found at the back of this section.

Table B-1. Equipment Required

Type

Specifications

Recommended Model

Digital multimeter

Resolution: 10nV @ 1V; Readout: 8 1/2

Agilent 3458A or

 

digits; Accuracy: 20 ppm

equivalent

Electronic load

20 V, 5A minimum, with transient capability

Agilent N3300A

 

and a a slew rate of 0.833A/∝s or better.

mainframe, with N3303A

 

 

module 6063A/B

GPIB controller

Full GPIB capabilities (only required if you

HP Series 200/300 or PC

 

are calibrating the unit over the GPIB)

with GPIB capability