5 - Programming the DC Source

Table 5-1. Bit Configurations of Status Registers

Bit

Signal

Meaning

 

 

 

Operation Status Group

 

0

CV

The selected output is in constant voltage mode (applies only in voltage priority mode)

1

CL+

The selected output is in positive current limit

(applies only in voltage priority mode)

2

CL-

The selected output is in negative current limit

(applies only in voltage priority mode)

3

CC

The selected output is in constant current mode (applies only in current priority mode)

4

VL+

The selected output is in positive voltage limit

(applies only in current priority mode)

5

VL-

The selected output is in negative voltage limit

(applies only in current priority mode)

6

OFF

The selected output is OFF

 

 

 

Questionable Status Group

 

0

OV+

The positive overvoltage protection has tripped

 

1

OV-

The negative overvoltage protection has tripped

2

PCLR

No communication with the selected output

 

4

OT

The overtemperature protection has tripped

 

10

UNR

The output is unregulated

 

12

OSC

The oscillation protection has tripped

 

14

MeasOvld

Output measurement exceeded capability of the range

 

 

Standard Event Status Group

 

0

OPC

Operation complete

 

2

QYE

Query error

 

3

DDE

Device-dependent error

 

4

EXE

Execution error

 

5

CME

Command error

 

7

PON

Power-on

 

 

 

Status Byte and Service Request Enable Registers

2

WTG

The unit is waiting for a trigger

 

3

QUES

Questionable status summary bit

 

4

MAV

Message Available summary bit

 

5

ESB

Event Status Summary bit

 

6

MSS

Master Status Summary bit

 

 

RQS

Request Service bit

 

7

OPER

Operation status summary bit

 

Operation Status Group

The Operation Status registers record signals that occur during normal operation. As shown below, the group consists of a Condition, PTR/NTR, Event, and Enable register. The outputs of the Operation Status register group are logically-ORed into the OPERation summary bit (7) of the Status Byte register.

Register

Command

Description

Condition

STAT:OPER:COND? (@<channel list>)

A register that holds real-time status of the circuits

 

 

being monitored. It is a read-only register.

PTR Filter

STAT:OPER:PTR <n> (@<channel list>)

A positive transition filter that functions as described

 

 

under STAT:OPER:NTRPTR commands in chapter 6.

 

 

It is a read/write register.

NTR Filter

STAT:OPER:NTR <n> (@<channel list>)

A negative transition filter that functions as described

 

 

under STAT:OPER:NTRPTR commands in chapter 6.

 

 

It is a read/write register.

Event

STAT:OPER:EVEN? (@<channel list>)

A register that latches any condition that is passed

 

 

through the PTR or NTR filters. It is a read-only

 

 

register that is cleared when read.

Enable

STAT:OPER:ENAB <n> (@<channel list>)

A register that functions as a mask for enabling specific

 

 

bits from the Event register. It is a read/write register.

50

Page 50
Image 50
Agilent Technologies N3280A manual Operation Status Group, Bit Configurations of Status Registers