4 Setup

4.1 Inserting the Optic

For a complete and working IVS, either the optic silent or the optic academy must be installed inside the CCD optic module.

A 1.5 mm allen key is used.

Do not touch optical surfaces.

All lenses must be clean.

Do not use force.

Turn the protecting cap at the end of the CCD optic module § counter clockwise and remove it to access the lens receptacle.

Remove the cover on the front side of the CCD optic module § by turning it counter-clockwise to access the alignment screw.

Make sure that the focus screw (S3) § and the alignment screws (S1, S2 and S4) § are in the middle of their range.

protecting cap

cover

S2

S3

F

S1 S4

Setup

13