Using MediCal Pro

Using MediCal Pro

Using MediCal Pro

After installation of the hardware and software, you must start MediCal Pro. With MediCal Pro, you can calibrate the display and projector, and start the projector stabilization.

Proceed as follows:

1.Start MediCal Pro. If appropriate, you can connect to MediCal Administrator.

2.During startup, MediCal Pro will detect the connected review display(s). They will be added automatically to the system configuration in MediCal Pro.

Note: In case the review displays were not automatically added to the configuration, you can add them manually, by running the Configuration Setup Wizard.

3.Add the projector(s) to the configuration by using the Configuration Setup wizard.

4.For each projector added to the configuration, MediCal Pro displays a dialog with which you can adjust the position and size of the patch for the optical sensor.

5.If you have not done this yet, install and connect the optical sensor, as described in the next paragraph. Or, you can position the patch over the already installed sensor.

6.After closing the sensor patch dialog (by clicking OK), MediCal Pro will check if a sensor is connected and installed properly.

7.For all the devices in the configuration, check if the DPMS setting is turned on. Therefore, right-click on the display icon and select Properties... from the drop-down menu. Then click on Details...

Check if the Powersave option is checked. If not, check it. This is necessary to use the DPMS possibilities of the imaging board.